48" CENTURY DOMESTIC HINGE
Report
- Report Number
- 2515984-2008-00009
- Event Type
- Other
- Date Received
- May 8, 2008
- Date of Event
- April 8, 2008
- Report Date
- May 6, 2008
- Manufacturer
- STERIS MEXICO S. DE R. L. DE C.V.
- Product Code
- FLE
- PMA / PMN Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NO INFORMATION
Narratives
THE MANUFACTURING ENGINEER ASSIGNED TO THE FAILURE INVESTIGATION, RULED OUT THAT THE DRAIN VALVE FAILURE. TWO ADDITIONAL COMPONENTS: THE STREAM TRAP (TR1) AND THE DRAIN SOLENOID VALVE (S40), MUST BOTH FAIL AT THE SAME TIME TO ACCUMULATE WATER IN THE CHAMBER. THEREFORE, THIS ISSUE IS LIKELY OPERATOR MISUSE. THE MANUAL OPERATOR (B) (4) CONTAINS THE WARNING SECTION (PAGE 1-2) THAT"...STEAM MAY BE RELEASED FROM THE CHAMBER WHEN DOOR IS OPENED TO MINIMIZE CONTACT WITH STEAM VAPOR". ADDITIONALLY, THE OPERATIVE INSTRUCTIONS (REF. MANUAL OPERATOR SECTION (B) (4) "DOOR UNSEALED) MAKES MENTION THE OPERATOR SHALL SILENCE THE ALARM AND CALL SERVICE.
THE HOSPITAL BIOMEDICAL TECHNICIAN WAS INVESTIGATING A REPORTED PROBLEM OF DOOR UNSEALED ALARM. WHEN SHE OPENED THE DOOR, A BURST OF STEAM CAME OUT AND HIT HER IN THE FACE. SHE COULD NOT BACK AWAY FROM THE STEAM BECAUSE THE LOADING CARTS WERE IN FRONT OF THE STERILIZER RESTRICTING HER MOVEMENTS. THE DOOR WAS UNSEALED AT THE TIME BUT CLOSED. THE FIELD SERVICE TECHNICIAN REPORTED A DRAIN CHECK VALVE HAD FAILED OPEN CAUSING WATER TO BE PULLED BACK UP INTO THE CHAMBER AND BOILED INTO STEAM. INJURIES WERE FIRST DEGREE ON FACE, NOSE, EYES AND MOUTH. THE BIOMEDICAL TECHNICIAN WAS TREATED AT THE ER AT THE FACILITY AND RETURNED TO WORK AFTERWARDS (SHE DID NOT TAKE ANYTIME OF WORK). EVIDENCE OF WARNING LABELS: (B) (4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 48" CENTURY DOMESTIC HINGE | STERILIZER | FLE | STERIS MEXICO S. DE R. L. DE C.V. | 1102321 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |