FDA Adverse Event Malfunction Summary report: N

HORIZON MEDICAL IMAGING

MDR report key: 1047865 · Received May 14, 2008

Report

Report Number
8022257-2008-00001
Event Type
Malfunction
Date Received
May 14, 2008
Date of Event
September 25, 2007
Report Date
May 13, 2008
Manufacturer
MCKESSON MEDICAL IMAGING COMPANY
Product Code
LLZ
PMA / PMN Number
K043146
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

MCKESSON MEDICAL IMAGING COMPANY ENGINEERING EXAMINATION REVEALED A SITUATION ON HORIZON MEDICAL IMAGING 11.0.X SOFTWARE ON WHICH THE IP LIBRARY HAS A REFRESH ERROR CAUSING ANNOTATION(S) MOVING IN THE VIEWPORT. THIS ERROR OCCURS ON AN IMAGE THAT HAS NON SQUARE PIXELS AND ANNOTATION(S), AND THE FIRST OPERATION APPLIED TO THAT IMAGE AFTER APPLYING A CUSTOM DISPLAY PROTOCOL IS ADJUST WINDOW/LEVEL. THE MCKESSON AFFECTED PRODUCTS ARE HORIZON RAD STATION ADVANCED OR HORIZON RAD STATION DISTRIBUTED VERSIONS 11.0.3 THROUGH 11.0.8 THAT IS PART OF HORIZON MEDICAL IMAGING 11.0.3 THROUGH 11.0.8 WORKSTATIONS. SOFTWARE UPDATES FOR HORIZON MEDICAL IMAGING 11.0.X WORKSTATIONS THAT ELIMINATE THE REPORTED PROBLEM WERE RELEASED IN 2008. SINCE THIS RELEASE DATE, ALL THE CLINICAL SITES WITH THE ABOVE LISTED PROGRAMS HAVE BEEN IN PROCESS OF BEING UPDATED WITH THE ABOVE LISTED PROGRAM UPDATES.

Description of Event or Problem · 1

THE ANNOTATIONS FOUND ON AN IMAGE WITH NON SQUARE PIXELS ARE SLIGHTLY MOVED IF THE FIRST OPERATION APPLIED TO THAT IMAGE AFTER APPLYING A CUSTOMER DISPLAY PROTOCOL IS ADJUST WINDOW/LEVEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HORIZON MEDICAL IMAGING PICTURE ARCHIVING AND COMMUNICATION SYS LLZ MCKESSON MEDICAL IMAGING COMPANY HMI11.0.X

Patients

Seq Age Sex Outcome Treatment
1