FDA Adverse Event Malfunction Summary report: N

HORIZON MEDICAL IMAGING

MDR report key: 1047320 · Received May 13, 2008

Report

Report Number
8022257-2008-00002
Event Type
Malfunction
Date Received
May 13, 2008
Date of Event
December 31, 2007
Report Date
May 13, 2008
Manufacturer
MCKESSON MEDICAL IMAGING COMPANY
Product Code
LMB
PMA / PMN Number
K043146
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MCKESSON MEDICAL IMAGING COMPANY ENGINEERING EXAMINATION REVEALED A RARE SITUATION ON HORIZON MEDICAL IMAGING 11.0.X SOFTWARE WITH HIERARCHICAL STORAGE MANAGEMENT ON WHICH TWO STUDIES SHARE THE SAME DESTINATION LOCATION, BAG AND SERIES. THE PROBLEM WAS CAUSED BY A HUMAN ERROR IN CONFIGURING THE ARCHIVE CONFIGURATION FILE MAKING TWO DEVICES SHARE THE SAME DESTINATION LOCATION. THE MCKESSON PRODUCTS THAT COULD BE AFFECTED ARE THE FOLLOWING HORIZON MEDICAL IMAGING VERSION: WITH HIERARCHICAL STORAGE MANAGEMENT (HSM). SINCE 2008, MCKESSON CUSTOMER SUPPORT HAS BEEN EXAMINING CUSTOMERS SITE'S CONFIGURATION SETTINGS AND, IF MISCONFIGURATION IS DETECTED, REVISING THEM TO ELIMINATE THE PROBLEM.

Description of Event or Problem · 1

WHEN ACCESSING MEDICAL IMAGES FOR A PARTICULAR PT, IMAGES FOR A DIFFERENT PT WERE DISPLAYED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HORIZON MEDICAL IMAGING PICTURE ARCHIVING AND COMMUNICATION SYS LMB MCKESSON MEDICAL IMAGING COMPANY HMI11.0.X

Patients

Seq Age Sex Outcome Treatment
1