HORIZON MEDICAL IMAGING
Report
- Report Number
- 8022257-2008-00002
- Event Type
- Malfunction
- Date Received
- May 13, 2008
- Date of Event
- December 31, 2007
- Report Date
- May 13, 2008
- Manufacturer
- MCKESSON MEDICAL IMAGING COMPANY
- Product Code
- LMB
- PMA / PMN Number
- K043146
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
MCKESSON MEDICAL IMAGING COMPANY ENGINEERING EXAMINATION REVEALED A RARE SITUATION ON HORIZON MEDICAL IMAGING 11.0.X SOFTWARE WITH HIERARCHICAL STORAGE MANAGEMENT ON WHICH TWO STUDIES SHARE THE SAME DESTINATION LOCATION, BAG AND SERIES. THE PROBLEM WAS CAUSED BY A HUMAN ERROR IN CONFIGURING THE ARCHIVE CONFIGURATION FILE MAKING TWO DEVICES SHARE THE SAME DESTINATION LOCATION. THE MCKESSON PRODUCTS THAT COULD BE AFFECTED ARE THE FOLLOWING HORIZON MEDICAL IMAGING VERSION: WITH HIERARCHICAL STORAGE MANAGEMENT (HSM). SINCE 2008, MCKESSON CUSTOMER SUPPORT HAS BEEN EXAMINING CUSTOMERS SITE'S CONFIGURATION SETTINGS AND, IF MISCONFIGURATION IS DETECTED, REVISING THEM TO ELIMINATE THE PROBLEM.
WHEN ACCESSING MEDICAL IMAGES FOR A PARTICULAR PT, IMAGES FOR A DIFFERENT PT WERE DISPLAYED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HORIZON MEDICAL IMAGING | PICTURE ARCHIVING AND COMMUNICATION SYS | LMB | MCKESSON MEDICAL IMAGING COMPANY | HMI11.0.X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |