ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAFETY LOCK
Report
- Report Number
- 1527736-1997-01614
- Event Type
- Malfunction
- Date Received
- July 10, 1997
- Report Date
- July 10, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
D6: DEVICE RETURNED WITH NO LOT ID. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY # 973795. VISUAL INSPECTIONS & RESULTS: LOCKOUT POSITION, FULLY FIRED; CARTRIDGE CONDITION, FULLY FIRED AND CARTRIDGE RETURN BATCH NUMBER J0088E. FUNCTIONAL TESTS & RESULTS: CONDITION OF FIRING TRIGGER LOCKOUT, GOOD; CONDITION OF PINION GEAR, GOOD; CONDITION OF SHORT RACK, GOOD; CONDITION OF YOKE, GOOD AND WAS INSTRUMENT CYCLED, NO. ANALYSIS CONCLUSION: BASED UPON THE INFO REC'D AND THE VISUAL EXAMINATION, IT WAS CONCLUDED THAT THE INSTRUMENT WAS RETURNED NON-FUNCTIONAL. THE INSTRUMENT WAS DISASSEMBLED AND FOUND TO HAVE A DAMAGED FIRING MECHANISM. NO CONCLUSION COULD BE REACHED AS TO HOW THIS DAMAGE OCCURRED. SOME CONDITIONS WHICH MIGHT RESULT IN DAMAGE TO THE FIRING MECHANISM ARE: INTERRUPTED FIRING CYCLE, TISSUE THICKER THAN INDICATED, ATTEMPTING TO FIRE THROUGH A SPENT CARTRIDGE, FIRING PRIOR TO COMPLETE CLAMPING OF THE JAWS AND FAILURE TO PROPERLY FOLLOW RELOADING INSTRUCTIONS. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE CO'S PRODUCTS.
IT WAS REPORTED THE DEVICE WAS USED DURING A MINI THORACOTOMY PROCEDURE WITH WEDGE RESECTION. IT WAS REPORTED THE EZ45G FIRED FINE TWICE. ON THE THIRD FIRING THE DEVICE WAS PLACED ACROSS LUNG TISSUE, BUT IT WOULD NOT FIRE. THERE WAS NO CONSEQUENCE TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAFETY LOCK | ENDOSCOPIC LINEAR CUTTER | GAG | ETHICON ENDO-SURGERY, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |