FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 1.0ML 31GA 6MM U40

MDR report key: 10463458 · Received August 27, 2020

Report

Report Number
1920898-2020-01123
Event Type
Malfunction
Date Received
August 27, 2020
Date of Event
August 8, 2020
Report Date
October 6, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 10/5/2020. H.6. INVESTIGATION: CUSTOMER RETURNED (19) 1CC, 6MM, 31G U40 INSULIN SYRINGES (9 IN AN OPEN POLY BAG, 10 IN A SEALED POLY BAG) WITH THE SHELF CARTON FROM LOT # 9231170. CUSTOMER STATES THAT THE SYRINGES ARE HARD TO INJECT, THEY ARE VERY TIGHT, AND THEY NEED A LOT OF ENERGY FOR INJECTING WHICH CAUSES PAIN, HARDSHIP AND AT TIMES BLEEDING. ALL RETURNED SYRINGES WERE TESTED (#1-9 FROM THE OPEN POLY BAG, #10-19 FROM THE SEALED POLY BAG) FOR POINT GEOMETRY, LUBE, AND CANNULA OD (SPECS: OUTER DIAMETER FOR 31G: 0.0100¿-0.0105¿). ALL OBSERVATIONS FALL WITHIN SPECIFICATIONS. ALSO, ALL SAMPLES WERE TESTED AND ALL WERE ABLE TO DRAW AND EXPEL PROPERLY WITHOUT ANY OBSERVED DEFECTS. ALL OBSERVATIONS FALL WITHIN SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9231170. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION [200842457] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. THERE WAS ONE (1) NOTIFICATION [200841002] NOTED FOR OUT OF SPEC BREAKOUT AND SUSTAINING. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PLUNGER IS DIFFICULT TO MOVE DURING INJECTION WITH 10 BD SYRINGE 1.0ML 31GA 6MM U40. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INSULIN SYRINGES THAT ARE HARD TO INJECT. I HAVE BEEN USING BD INSULIN SYRINGES 31G FOR LONG TIME. BUT OF LATE I AM FINDING QUALITY HAS COME DOWN. THEY ARE VERY TIGHT AND HARD.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PLUNGER IS DIFFICULT TO MOVE DURING INJECTION WITH 10 BD SYRINGE 1.0ML 31GA 6MM U40. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INSULIN SYRINGES THAT ARE HARD TO INJECT. I HAVE BEEN USING BD INSULIN SYRINGES 31G FOR LONG TIME. BUT OF LATE I AM FINDING QUALITY HAS COME DOWN. THEY ARE VERY TIGHT AND HARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
928190 BD SYRINGE 1.0ML 31GA 6MM U40 PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 9231170

Patients

Seq Age Sex Outcome Treatment
1 Other