STYLE 120 SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2020-13945
- Event Type
- Injury
- Date Received
- August 26, 2020
- Date of Event
- July 30, 2020
- Report Date
- October 2, 2020
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- UDI-DI
- 10888628002548
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5,D7,H6. VISUAL EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED DEVICE PATCH WITH LOT NUMBER 2672621, CATALOG NUMBER 120-340, RED PARTICLES ON THE SHELL, RED BIOLOGICAL TISSUE AND AN OPENING EXTENDED. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE.
DEVICE HAS BEEN EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE "RUPTURE." DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 920712 | STYLE 120 SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 120-340 | 2672621 | 10888628002548 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Required Intervention |