SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP
Report
- Report Number
- 1920898-2020-01112
- Event Type
- Malfunction
- Date Received
- August 25, 2020
- Date of Event
- August 4, 2020
- Report Date
- September 8, 2020
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: CUSTOMER RETURNED (1) LOOSE 3/10CC, 12.7MM SYRINGE. CUSTOMER STATES THAT WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED, LIKE CUT OFF, FROM THE HUB. THE RETURNED SYRINGE WAS TESTED AND THE HUB-NEEDLE ASSEMBLY DID NOT SEPARATE FROM THE BARREL WHEN THE SHIELD WAS REMOVED. UNABLE TO PERFORM DHR CHECK FOR NEEDLE HUB SEPARATES DUE TO UNKNOWN LOT NUMBER. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT THE SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP EXPERIENCED HUB SEPARATION FROM THE DEVICE PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED, LIKE CUT OFF, FROM THE HUB.
DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).
IT WAS REPORTED THAT THE SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP EXPERIENCED HUB SEPARATION FROM THE DEVICE PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED, LIKE CUT OFF, FROM THE HUB.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 916596 | SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP | SYRINGE | FMF | BD MEDICAL - DIABETES CARE | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |