FDA Adverse Event Injury Summary report: N

LOGIQ E9 BREAKTHROUGH 2011

MDR report key: 10438127 · Received August 21, 2020

Report

Report Number
3005860720-2020-00004
Event Type
Injury
Date Received
August 21, 2020
Date of Event
July 8, 2020
Report Date
September 24, 2020
Manufacturer
GE MEDICAL SYSTEMS ULTRASOUND &
Product Code
IYN
PMA / PMN Number
K163077
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

GEHC'S INVESTIGATION HAS COMPLETED. THE EVENT IS AN ALLEGATION BY THE CUSTOMER OF EXACERBATION OF AN EXISTING MSK INJURY (KNEE INJURY) OF A SONOGRAPHER WHO MOVED THE LOGIQ E9 SCANNER WHILE TURNING A CORNER. GEHC EVALUATED THE LOGIQ E9 AND FOUND NO MALFUNCTION OF THE DEVICE. ADDITIONALLY, THE COMPLAINT FILE WAS SEARCHED FOR OTHER REPORTS OF THE SAME AND/OR SIMILAR EVENTS AND NO OTHER RECORDS WERE FOUND. THE RISK EVALUATION CONCLUDED PROPER MITIGATIONS ARE IN PLACE, AND NO FURTHER ACTION IS BEING PERFORMED.

Additional Manufacturer Narrative · 1

NO REPORT OF PATIENT INVOLVEMENT, THIS IS A USER INJURY. LEGAL MANUFACTURER NAME IS GE MEDICAL SYSTEMS ULTRASOUND & PRIMARY CARE DIAGNOSTICS LLC. A GEHC SERVICE REPRESENTATIVE EVALUATED THE EQUIPMENT AND GEHC'S INVESTIGATION IS ONGOING. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Description of Event or Problem · 1

A GEHC SERVICE REPRESENTATIVE WAS MADE AWARE FROM A CUSTOMER THAT AN ULTRASOUND TECHNICIAN INJURED THEIR KNEE WHILE MOVING THE PORTABLE ULTRASOUND SYSTEM, THE LOGIQ E9. THE CUSTOMER PROVIDED ADDITIONAL INFORMATION STATING THE INJURY OCCURRED AFTER THE TECHNICIAN PERFORMED A PORTABLE EXAM ON THE LABOR AND DELIVERY UNIT. THE EMPLOYEE STATED THAT THEY WERE TRANSPORTING THE LOGIQ E9 FROM THE MONITOR HANDLES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
902800 LOGIQ E9 BREAKTHROUGH 2011 SYSTEM, IMAGING, PULSED DOPPLER, ULT IYN GE MEDICAL SYSTEMS ULTRASOUND & UGLG9L 00000123788US2

Patients

Seq Age Sex Outcome Treatment
1 Other