CARPENTIER-EDWARDS SUPRAANNULAR BIOPROSTHESIS
Report
- Report Number
- 2015691-2020-13250
- Event Type
- Injury
- Date Received
- August 21, 2020
- Date of Event
- July 28, 2020
- Report Date
- July 28, 2020
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- DYE
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS DEVICE IS NOT SOLD OR MARKETED IN THE UNITED STATES; HOWEVER, IT IS SIMILAR TO THE BRAND CARPENTIER-EDWARDS DURAFLEX LOW PRESSURE MITRAL BIOPROSTHESIS, MODEL# 6625-LP, PMA# P870077. TISSUE DEGENERATION RELATED STRUCTURAL DETERIORATION EITHER CALCIFIC OR NON-CALCIFIC ARE COMMON CHRONIC FAILURE MODES FOR THIS TYPE OF BIOPROSTHETIC HEART VALVES. THE OPERATIONAL MECHANICAL STRESS AND BIOLOGICAL FACTORS ARE GENERALLY BELIEVED TO BE THE MAJOR CONTRIBUTORS TO THE NON-CALCIFIC BIOPROSTHETIC TISSUE DEGENERATION. STRUCTURAL VALVE DETERIORATION (SVD) CAN, AND TYPICALLY DOES, LEAD TO CHRONIC CENTRAL LEAKS OVER A PERIOD OF TIME. SVD IS THE MOST COMMON REASON FOR BIOPROSTHESIS EXPLANT AND ENCOMPASSES MULTIPLE FAILURE MODES, INCLUDING CALCIFICATION, NONCALCIFIC DEGENERATION, DEHISCENCE, CUSP THICKENING OR FIBROSIS, OR A COMBINATION OF THESE. SUCH FAILURE MODES MAY OCCUR SINGULARLY OR CONCOMITANTLY. THE DEVICE WAS NOT RETURNED FOR EVALUATION, AS IT REMAINS IMPLANTED. THE ROOT CAUSE OF THIS EVENT CANNOT BE CONCLUSIVELY DETERMINED. HOWEVER, IT IS LIKELY THAT PATIENT RELATED FACTORS AND THE PROGRESSION OF THE UNDERLYING VALVULAR DISEASE PATHOLOGY CONTRIBUTED TO THE EVENT. THE DEVICE HISTORY RECORD (DHR) WAS NOT ABLE TO BE REVIEWED AS THE DEVICE SERIAL NUMBER WAS NOT PROVIDED. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL EVENTS. TRENDS ARE MONITORED ON A MONTHLY BASIS AND IF ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED.
EDWARDS RECEIVED NOTIFICATION THAT A 33MM VALVE IMPLANTED IN AN UNKNOWN POSITION WAS DISABLED VIA A VALVE-IN-VALVE REPLACEMENT AFTER AN APPROXIMATE IMPLANT DURATION OF 17 YEARS AND 6 MONTHS DUE TO DEGENERATION LEADING TO STENOSIS AND REGURGITATION. A 29MM TRANSCATHETER VALVE WAS IMPLANTED WITH GOOD RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 900116 | CARPENTIER-EDWARDS SUPRAANNULAR BIOPROSTHESIS | HEART-VALVE, REPLACEMENT | DYE | EDWARDS LIFESCIENCES | 6650 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Hospitalization| R |