CANGAROO ENVELOPE
Report
- Report Number
- 3005619880-2020-00041
- Event Type
- Injury
- Date Received
- August 21, 2020
- Date of Event
- July 23, 2020
- Report Date
- August 21, 2020
- Manufacturer
- AZIYO BIOLOGICS INC.
- Product Code
- FTM
- UDI-DI
- 00859389005034
- PMA / PMN Number
- K192616
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
MANUFACTURING REVIEW OF THE CANGAROO ENVELOPE DEVICE HISTORY RECORD FOR THE REPORTED LOT SHOWS THAT ALL PACKAGED AND LABELED UNITS WERE QUALITY RELEASED TO DISTRIBUTION ON 03/03/2020 HAVING MET ALL INTERNAL QC ACCEPTANCE REQUIREMENTS. ALL STERILIZATION PROCESSING RECORDS AND BIOBURDEN TESTING INDICATE A SUCCESSFUL STERILIZATION CYCLE AND PASSING RESULTS OF PRODUCT LAL AND STERILITY SAMPLES ALLOWING THE SUBASSEMBLY LOT TO BE RELEASED FOR FURTHER PACKAGING AND LABELING HAVING MET ALL CRITERIA FOR RELEASE. THERE WERE NO NON-CONFORMANCES DURING MANUFACTURING OR STERILIZATION POTENTIALLY IMPACTING THE FINAL ACCEPTANCE OF THIS MANUFACTURING LOT. IN LIEU OF A REQUEST FOR THE OEM SUPPLIER DHR REVIEW, IT IS NOTED THAT AZIYO PROCESSES THE NON-STERILE ENVELOPE MATERIALS INCLUDING CUTTING, SUTURING, PACKAGING AND STERILIZING THE PRODUCT. IT IS ALSO NOTED THAT PER THE INSTRUCTIONS FOR USE (IFU - ART-20662) PROVIDED WITH THE FINISHED CANGAROO ENVELOPE DEVICE, INFECTION IS LISTED AS A POTENTIAL COMPLICATION ASSOCIATED WITH THE PROCEDURE AND DEVICE. ALTHOUGH THE EXACT CAUSE OF THE REPORTED INFECTION CANNOT BE CONCLUSIVELY DETERMINED, INFECTION IS A KNOWN COMPLICATION ASSOCIATED WITH THE USE OF A CANGAROO ENVELOPE AND A SURGICAL IMPLANT PROCEDURE. SHOULD AZIYO RECEIVE ANY ADDITIONAL DETAILS RELATED TO THIS EVENT, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED BY AZIYO BIOLOGICS' BUSINESS PARTNER BOSTON SCIENTIFIC, THAT A CANGAROO ENVELOPE (MODEL CMCV-009-MED LOT# M20C1091) WAS USED WITH A BOSTON SCIENTIFIC DEVICE (MODEL #/SERIAL # UNKNOWN) AND IMPLANTED ON (B)(6) 2020. PATIENT DEVELOPED AN INFECTION AND ON (B)(6) 2020 THE ENTIRE SYSTEM WAS EXPLANTED. A NEW SYSTEM WAS IMPLANTED AT THE SAME TIME ON THE OPPOSITE SIDE OF PATIENT'S CHEST. THE PATIENT WAS TREATED WITH INTRAVENOUS ANTIBIOTICS AND HAS SINCE RECOVERED. THE ISSUE WAS REPORTED TO AZIYO BIOLOGICS ON 7/24/2020 WITH ADDITIONAL DETAIL PROVIDED ON 8/05/2020.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 902678 | CANGAROO ENVELOPE | MESH, SURGICAL - FTM | FTM | AZIYO BIOLOGICS INC. | CMCV-009-MED | M20C1091 | 00859389005034 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | BSC CIED (MODEL# / SERIAL# - UNKNOWN) |