FDA Adverse Event
Malfunction
Summary report: N
HEATER-COOLER SYSTEM 3T
MDR report key: 10435822
·
Received August 21, 2020
Report
- Report Number
- 9611109-2020-00476
- Event Type
- Malfunction
- Date Received
- August 21, 2020
- Date of Event
- July 24, 2020
- Report Date
- July 24, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND GMBH
- Product Code
- DWC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO KNOWN PATIENT INVOLVEMENT. PMA/510(K). THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). FIELD TECHNICAL ENGINEER REPLACED PUMP PATIENT 2 AND DISTRIBUTION BOARD AND PERFORMED FUNCTIONAL TEST WITH POSITIVE RESULTS. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
Description of Event or Problem · 1
LIVANOVA RECEIVED A REPORT ABOUT HEATER COOLER 3T ALARMING E22. THERE IS NO EVIDENCE OF ANY PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 899955 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND GMBH | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |