AUTOSTAINER LINK 48
Report
- Report Number
- 2022180-2020-00255
- Event Type
- Malfunction
- Date Received
- August 20, 2020
- Date of Event
- July 10, 2018
- Report Date
- August 20, 2020
- Manufacturer
- DAKO NORTH AMERICA, INC.
- Product Code
- KPA
- UDI-DI
- 05700572035497
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AS PART OF AGILENT'S CONTINUOUS IMPROVEMENT PROCESS, OUR PROCEDURES WERE REVIEWED AND IT WAS DETERMINED THAT THERE IS A NEED TO REPORT POTENTIAL RISKS ASSOCIATED WITH CRITICAL LIQUID DISPENSE SYSTEM COMPONENTS ON THE AUTOSTAINER. WE HAVE ASSESSED ALL AUTOSTAINER DISPENSE AND LEAKAGE COMPLAINTS FROM NOVEMBER OF 2017 THROUGH MARCH OF 2020 THAT RESULTED IN ALTERATION IN STAINING OR POTENTIAL ALTERATION IN SLIDE STAINING. FOLLOWING THIS REVIEW, IT WAS DECIDED TO RETROSPECTIVELY FILE AN MDR FOR CUSTOMER COMPLAINT (B)(4). THIS IS AN INITIAL REPORT. BASED ON COMPLAINT REPORT OR INVESTIGATED FAILURE MODE, THERE WAS POTENTIAL FOR A STAINING ALTERATION. CUSTOMER COMPLAINT RECORD REPORTED THE EVENT AS FOLLOWS: BUFFER TUBE FILLED WITH AIR. NO DIRECT OR INDIRECT PATIENT HARM OR USER HARM HAS BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 899644 | AUTOSTAINER LINK 48 | AUTOSTAINER LINK 48 INSTRUMENT SYSTEM | KPA | DAKO NORTH AMERICA, INC. | AS480 | 05700572035497 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |