FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 10433080 · Received August 20, 2020

Report

Report Number
9617229-2020-12799
Event Type
Injury
Date Received
August 20, 2020
Date of Event
July 1, 2020
Report Date
September 22, 2020
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
10888628007307
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

VISUAL EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED DEVICE PATCH WITH LOT NUMBER 3181809, CATALOG NUMBER SRF-345 AND DEFORMATION. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE.

Description of Event or Problem · 0

DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE IV.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
898955 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) SRF-345 3181809 10888628007307

Patients

Seq Age Sex Outcome Treatment
1 34 YR Required Intervention