SYRINGE 1ML LL W/O DN
Report
- Report Number
- 1213809-2020-00555
- Event Type
- Malfunction
- Date Received
- August 20, 2020
- Date of Event
- July 29, 2020
- Report Date
- September 15, 2020
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- UDI-DI
- 30382903096283
- PMA / PMN Number
- K941562
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10 DEVICE AVAILABLE FOR EVAL YES. D.10 RETURNED TO MANUFACTURER ON: 2020-08-26. H.6. INVESTIGATION SUMMARY: SIX 1ML SYRINGES IN BLISTER PACKS FROM BATCH: 8001745 (P/N: 309628) WERE RECEIVED AND EVALUATED. FIVE PACKAGES WERE FULLY SEALED, AND ONE WAS OPENED. ALL THE SYRINGES APPEARED TO BE PROPERLY ASSEMBLED WITH ALL COMPONENTS PRESENT. NO DEFECT WAS OBSERVED. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.
IT WAS REPORTED THAT SYRINGE 1ML LL W/O DN HAD A MOLDING DEFECT, SHORT SHOTS ON THE BARREL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "RECENTLY WE HAVE NOTICED THAT THE "STAMP STOP" (RESISTANCE THAT PREVENTS THE SYRINGE STAMP FROM BEING PULLED OUT) IS MISSING."
(B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT SYRINGE 1ML LL W/O DN HAD A MOLDING DEFECT, SHORT SHOTS ON THE BARREL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "RECENTLY WE HAVE NOTICED THAT THE "STAMP STOP" (= RESISTANCE THAT PREVENTS THE SYRINGE STAMP FROM BEING PULLED OUT) IS MISSING."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 899451 | SYRINGE 1ML LL W/O DN | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 309628 | 9260035 | 30382903096283 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |