FDA Adverse Event Malfunction Summary report: N

TRILOGY EVO

MDR report key: 10424800 · Received August 19, 2020

Report

Report Number
2518422-2020-01920
Event Type
Malfunction
Date Received
August 19, 2020
Date of Event
August 7, 2020
Report Date
August 7, 2020
Manufacturer
RESPIRONICS, INC
Product Code
CBK
PMA / PMN Number
K181166
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE MANUFACTURER PREVIOUSLY REPORTED THE INCORRECT 510K NUMBER AS K18166. THE CORRECT 510K NUMBER IS K181166 AND IS REFLECTED IN G5.

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THE DEVICE WAS NOT IN PATIENT USE. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S INTERNAL BATTERY WAS REPLACED TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
890507 TRILOGY EVO VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC DS2110X11B

Patients

Seq Age Sex Outcome Treatment
1