FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 1042427 · Received May 8, 2008

Report

Report Number
2124215-2008-34214
Event Type
Malfunction
Date Received
May 8, 2008
Date of Event
March 7, 2008
Report Date
September 22, 2017
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
LAY USER/PATIENT

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THE DEVICE WAS LATER EXPLANTED AND REPLACED 8 YEARS LATER FOR REPORTED NORMAL BATTERY DEPLETION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSIGNIA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND 1297 NA

Patients

Seq Age Sex Outcome Treatment
1 85 YR Other| O