FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK TIP SYRINGE

MDR report key: 10418283 · Received August 17, 2020

Report

Report Number
1213809-2020-00539
Event Type
Malfunction
Date Received
August 17, 2020
Date of Event
July 27, 2020
Report Date
August 19, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096573
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: ONE PHOTO OF A 3ML SYRINGE IN A FULLY SEALED BLISTER PACK WAS RECEIVED AND EVALUATED. IT WAS OBSERVED THERE WAS NO VISIBLE PRINT ON THE SYRINGE, WHICH WAS REJECTABLE PER PRODUCT SPECIFICATION. A POTENTIAL ROOT CAUSE FOR THE MISSING SCALE DEFECT IS ASSOCIATED WITH THE MARKING PROCESS. THE AQL FOR MISSING PRINT IS 0.25%. (B)(4). NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. BATCH 0150246 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ TIP SYRINGE HAD NO SCALE MARKINGS ON IT PRIOR TO USE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE GOT A REPORT EARLIER TODAY THAT THE FLOORS HAVE BEEN FINDING, OVER THE LAST COUPLE OF WEEKS, THAT SOME OF THE 3CC SYRINGES DON'T HAVE ANY VOLUME MARKINGS ON THEM. THE ONE THEY FOUND WAS LOT# 0150246."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD LUER-LOK¿ TIP SYRINGE HAD NO SCALE MARKINGS ON IT PRIOR TO USE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE GOT A REPORT EARLIER TODAY THAT THE FLOORS HAVE BEEN FINDING, OVER THE LAST COUPLE OF WEEKS, THAT SOME OF THE 3CC SYRINGES DON'T HAVE ANY VOLUME MARKINGS ON THEM. THE ONE THEY FOUND WAS LOT# 0150246."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
876564 BD LUER-LOK TIP SYRINGE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309657 0150246 30382903096573

Patients

Seq Age Sex Outcome Treatment
1 Other