FDA Adverse Event Injury Summary report: N

INSIGNIA I

MDR report key: 1041569 · Received May 8, 2008

Report

Report Number
2124215-2008-31958
Event Type
Injury
Date Received
May 8, 2008
Date of Event
May 1, 2007
Report Date
February 5, 2020
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
LWP
PMA / PMN Number
N970003/S024
Removal / Correction Number
Z1293/99-6
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING, AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. LABORATORY ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 1

THIS DEVICE WAS LATER EXPLANTED DUE TO ROUTINE CHANGE OUT. NO ADVERSE PATIENT EFFECTS WERE REPORTED RELATED TO THE EXPLANT PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSIGNIA I IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) LWP BOSTON SCIENTIFIC CORPORATION 1291 122639

Patients

Seq Age Sex Outcome Treatment
1 58 YR Other| R| R