FDA Adverse Event Malfunction Summary report: N

PHOENIX

MDR report key: 10413295 · Received August 17, 2020

Report

Report Number
10413295
Event Type
Malfunction
Date Received
August 17, 2020
Date of Event
May 21, 2020
Report Date
June 11, 2020
Manufacturer
GAMBRO DASCO SPA
Product Code
KDI
Product Problem
Yes
Report Source
User Facility report
Reporter Location
ND
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

AFTER INSTALLING MY NEW STERILANT TANK ON THE PHOENIX UNIT THIS MORNING. I PROCEEDED TO TRY AND CALIBRATE THE B&A CONDUCTIVITY ON THE UNIT. I DID THIS BECAUSE, THIS UNIT HAS BEEN HAVING ALL KINDS OF INTERMITTENT CONDUCTIVITY ERRORS. DURING THE CALIBRATION, THE UNIT KEPT ON TRIPPING ERRORS, THAT DIALYZER PRESSURES WERE HIGH AND LOW. AFTER INSPECTING THE PI/PD/PFS PRESSURE TRANSDUCERS, I DISCOVERED THAT THE DIAPHRAGM IN ALL THREE OF THEM WAS DAMAGED/DETERIORATED RUBBER. THIS IS NOT THE FIRST UNIT THAT I HAVE FOUND THIS PROBLEM ON; THIS IS A WIDE SPREAD PROBLEM. BECAUSE OF THIS DISCOVERY I WILL NEED TO CHECK EVERY SMALL ORIFICE FOR CLOGGING. THE PART # 6990717. I HAD ANOTHER PRESSURE TRANSDUCER GO BAD ON ANOTHER UNIT THIS MORNING WHILE DOING A PH1 INSPECTION THIS MORNING. I WAS UNABLE TO TEST/CALIBRATE PI THIS MORNING. WHEN I OPENED UP THE PRESSURE TRANSDUCER IT WAS THE SAME WAY. ADDITIONALLY I DID ANOTHER INSPECTION ON ANOTHER UNIT ON MONDAY THAT WAS THE SAME WAY ON PO. LATELY EVERY UNIT THAT I AM TOUCHING HAS BAD TRANSDUCERS. MANUFACTURER RESPONSE FOR HEMODIALYSIS UNIT #3, GAMBRO (PHOENIX) (PER SITE REPORTER). THANK YOU FOR CONTACTING BAXTER AND REPORTING ABOUT THE PRODUCT COMPLAINT PRESSURE TRANSDUCERS WITH PRODUCT CODE: 6990717, WHEREIN, AFTER INSPECTING THE PI/PD/PFS PRESSURE TRANSDUCERS, IT WAS DISCOVERED THAT THE DIAPHRAGM IN ALL THREE OF THEM WERE DAMAGED/DETERIORATED RUBBER. HAD ENDED FINDING THE FOURTH PRESSURE TRANSDUCER DIAPHRAGM THAT WAS ALSO DETERIORATING AND POSSIBLY BLOCKING OTHER ORIFICES IN THE HYDRAULIC SYSTEM. WE WOULD LIKE TO REQUEST ADDITIONAL INFORMATION RELATED TO YOUR REPORT AND HOPING YOU CAN ASSIST US WITH THIS REQUEST. WOULD YOU BE ABLE TO PROVIDE ASSISTANCE IN OBTAINING THE FOLLOWING INFORMATION: (THIS IS FOR SECOND FOLLOW-UP) ARE THE ACTUAL SAMPLES AVAILABLE TO BE RETURNED TO MANUFACTURING PLANT FOR FURTHER INVESTIGATION? WE REGRET ANY INCONVENIENCE THIS ISSUE MAY HAVE CAUSED AND THANK YOU FOR BRINGING THIS MATTER TO OUR ATTENTION. BAXTER HEALTHCARE IS COMMITTED TO TREATING CUSTOMER COMPLAINTS SERIOUSLY AND WE MONITOR THEM CLOSELY. WE APPRECIATE YOU TAKING THE TIME AND THANK YOU IN ADVANCE FOR YOUR ASSISTANCE. MANUFACTURER RESPONSE FOR HEMODIALYSIS UNIT, GAMBRO (PHOENIX) (PER SITE REPORTER). THANK YOU FOR CONTACTING BAXTER AND REPORTING ABOUT THE PRODUCT COMPLAINT PRESSURE TRANSDUCERS WITH PRODUCT CODE: 6990717, WHEREIN, AFTER INSPECTING THE PI/PD/PFS PRESSURE TRANSDUCERS, IT WAS DISCOVERED THAT THE DIAPHRAGM IN ALL THREE OF THEM WERE DAMAGED/DETERIORATED RUBBER. HAD ENDED FINDING THE FOURTH PRESSURE TRANSDUCER DIAPHRAGM THAT WAS ALSO DETERIORATING AND POSSIBLY BLOCKING OTHER ORIFICES IN THE HYDRAULIC SYSTEM. WE WOULD LIKE TO REQUEST ADDITIONAL INFORMATION RELATED TO YOUR REPORT AND HOPING YOU CAN ASSIST US WITH THIS REQUEST. WOULD YOU BE ABLE TO PROVIDE ASSISTANCE IN OBTAINING THE FOLLOWING INFORMATION: (THIS IS FOR SECOND FOLLOW-UP). ARE THE ACTUAL SAMPLES AVAILABLE TO BE RETURNED TO MANUFACTURING PLANT FOR FURTHER INVESTIGATION? WE REGRET ANY INCONVENIENCE THIS ISSUE MAY HAVE CAUSED AND THANK YOU FOR BRINGING THIS MATTER TO OUR ATTENTION. BAXTER HEALTHCARE IS COMMITTED TO TREATING CUSTOMER COMPLAINTS SERIOUSLY AND WE MONITOR THEM CLOSELY. WE APPRECIATE YOU TAKING THE TIME AND THANK YOU IN ADVANCE FOR YOUR ASSISTANCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
880782 PHOENIX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM KDI GAMBRO DASCO SPA PHOENIX 6990717

Patients

Seq Age Sex Outcome Treatment
1