LARGE NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2008-00259
- Event Type
- Malfunction
- Date Received
- May 9, 2008
- Report Date
- May 9, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- GCJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING FOUND THE INSTRUMENT HAD ZERO USES REMAINING, THEREFORE, WAS UNABLE TO CONDUCT PERFORMANCE TESTS ON A SYSTEM. MANUALLY ROTATING THE INPUT DISCS CONFIRMED THE INSTRUMENT MOVES INTUITIVELY WITH FULL RANGE OF MOTION, AND THE GRIPS OPEN AND CLOSE PROPERLY. ENGINEERING ALSO OBSERVED THE DISTAL END OF MAIN TUBE HAS VARIOUS DEEP SCRATCHES WITH MATERIAL REMOVED. SCRATCHES ARE NOT ALIGNED WITH THE TUBE AXIS, SUGGESTING THEY WERE CAUSED BY INSTRUMENT COLLISIONS. INSTRUMENT IS FULLY FUNCTIONAL OTHERWISE. NO OTHER DAMAGE FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT DURING A DA VINCI S SURGICAL PROCEDURE, A LARGE NEEDLE DRIVER INSTRUMENT WAS NOT HOLDING THE NEEDLE. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED, AND NO PIECES FELL INTO THE PT. NO PT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LARGE NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | GCJ | INTUITIVE SURGICAL, INC. | 420006-05 | 0712102 369 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESSORIES| DA VINCI S SURGICAL SYSTEM |