FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 1041296 · Received May 9, 2008

Report

Report Number
2955842-2008-00256
Event Type
Malfunction
Date Received
May 9, 2008
Date of Event
April 9, 2008
Report Date
May 9, 2008
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING FOUND ONE SCISSOR BLADE BROKE OFF AT THE CROSS SECTION OF THE GRIP CABLE CRIMP. THE BROKEN BLADE WAS RETURNED. THE INTACT BLADE SHOWS SIGNS OF LIGHT PITTING ON THE NON-CUTTING EDGE, INDICATING POSSIBLE CORROSION WHICH MAY HAVE CAUSED CORROSIVE CRACKING, CAUSING THE BLADE FAILURE. LABORATORY ANALYSIS DISCOVERED CHLORINE PRESENT ON THE FRACTURE SURFACE, INDICATING THAT THE TIP BREAKAGE WAS DUE TO STRESS CORROSION CRACKING. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL INSTRUCTIONS, CLEANING THE INSTRUMENTS. DO NOT EXPOSE INSTRUMENTS TO HYDROGEN PEROXIDE (H202), BLEACH OR ALKALINE-BASED CLEANING AGENTS, AS THIS MAY RESULT IN INSTRUMENT DAMAGE. PROLONGED EXPOSURE TO EITHER ULTRASONIC CLEANING OR CLEANING AGENTS MAY RESULT IN INSTRUMENT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S PROSTATECTOMY PROCEDURE, WHILE THE SURGEON WAS ATTEMPTING TO CUT TISSUE WITH THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT, ONE HALF OF THE TIP BROKE OFF AND FELL INTO THE PATIENT. THE FRAGMENT WAS IMMEDIATELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER INSTRUMENT OF THE SAME TYPE, WITH APPROXIMATELY A 10 MINUTE DELAY. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONOPOLAR CURVED SCISSORS INSTRUMENT ELECTROSURGICAL INSTRUMENT GEI INTUITIVE SURGICAL, INC. 420179-08 1406071 087

Patients

Seq Age Sex Outcome Treatment
1 ACCESSORIES| ELECTROSURGICAL UNIT| DA VINCI S SURGICAL SYSTEMS