MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2008-00256
- Event Type
- Malfunction
- Date Received
- May 9, 2008
- Date of Event
- April 9, 2008
- Report Date
- May 9, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING FOUND ONE SCISSOR BLADE BROKE OFF AT THE CROSS SECTION OF THE GRIP CABLE CRIMP. THE BROKEN BLADE WAS RETURNED. THE INTACT BLADE SHOWS SIGNS OF LIGHT PITTING ON THE NON-CUTTING EDGE, INDICATING POSSIBLE CORROSION WHICH MAY HAVE CAUSED CORROSIVE CRACKING, CAUSING THE BLADE FAILURE. LABORATORY ANALYSIS DISCOVERED CHLORINE PRESENT ON THE FRACTURE SURFACE, INDICATING THAT THE TIP BREAKAGE WAS DUE TO STRESS CORROSION CRACKING. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL INSTRUCTIONS, CLEANING THE INSTRUMENTS. DO NOT EXPOSE INSTRUMENTS TO HYDROGEN PEROXIDE (H202), BLEACH OR ALKALINE-BASED CLEANING AGENTS, AS THIS MAY RESULT IN INSTRUMENT DAMAGE. PROLONGED EXPOSURE TO EITHER ULTRASONIC CLEANING OR CLEANING AGENTS MAY RESULT IN INSTRUMENT DAMAGE.
IT WAS REPORTED THAT DURING A DA VINCI S PROSTATECTOMY PROCEDURE, WHILE THE SURGEON WAS ATTEMPTING TO CUT TISSUE WITH THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT, ONE HALF OF THE TIP BROKE OFF AND FELL INTO THE PATIENT. THE FRAGMENT WAS IMMEDIATELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER INSTRUMENT OF THE SAME TYPE, WITH APPROXIMATELY A 10 MINUTE DELAY. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ELECTROSURGICAL INSTRUMENT | GEI | INTUITIVE SURGICAL, INC. | 420179-08 | 1406071 087 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESSORIES| ELECTROSURGICAL UNIT| DA VINCI S SURGICAL SYSTEMS |