FDA Adverse Event Malfunction Summary report: N

STERRAD 200 STERILIZATION SYSTEM

MDR report key: 1040854 · Received May 9, 2008

Report

Report Number
2084725-2008-00164
Event Type
Malfunction
Date Received
May 9, 2008
Date of Event
April 18, 2008
Report Date
April 18, 2008
Manufacturer
YAMATO LAB-TECH CO., LTD
Product Code
MLR
PMA / PMN Number
K030429
Removal / Correction Number
Z-1630-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CAPITAL EQUIPMENT, FSE WENT TO THE FACILITY. THE FSE REPLACED THE OIL MIST FILTER AND THE CATALYTIC CONVERTER BECAUSE THERE WAS "OIL MIST" COMING OUT OF IT. HE VERIFIED THAT NO OIL MIST WAS COMING OUT OF IT. HE VERIFIED THAT THE SYSTEM MET MFR SPECIFICATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A BURNING SMELL COMING FROM THE UNIT WHEN IT WAS RUNNING. THEY STATED THAT THE SMELL WOULD OCCUR DURING THE "INJECTION PHASE". THE UNIT WAS NOT CANCELING. THE CUSTOMER STATED THAT THERE WERE NO PHYSICAL COMPLAINTS REPORTED. THE ASP FIELD SERVICE ENGINEER (FSE) WENT TO THE FACILITY TO REPAIR THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERRAD 200 STERILIZATION SYSTEM STERILIZER MLR YAMATO LAB-TECH CO., LTD 10201 NA

Patients

Seq Age Sex Outcome Treatment
1