FDA Adverse Event Malfunction Summary report: N

COSYCOT INFANT WARMER

MDR report key: 1040835 · Received May 9, 2008

Report

Report Number
9611451-2008-00285
Event Type
Malfunction
Date Received
May 9, 2008
Report Date
March 14, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
FMT
PMA / PMN Number
K971695
Removal / Correction Number
Z-0485-2007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL MANUFACTURING DATES: 10/13/2005, 10/13/2005. REPLACED THE CRACKED BASES OF COSYCOT INFANT WARMERS. AN INVESTIGATION WAS CARRIED OUT BASED ON THE EVENT DESCRIPTION AND RESULTS OF PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS, AS THE COMPLAINT DEVICES HAVE NOT BEEN RETURNED FOR EVALUATION. CRACKS ON THE ELEVATOR BASE OF THE COSYCOT INFANT WARMER, DUE TO EXCESSIVE WEIGHT LOADING, IS A KNOWN PROBLEM. TOTAL WEIGHT OF THE DEVICE, INCLUDING ACCESSORIES AND PATIENT, EXCEEDING 115 KG MAY, UNDER CERTAIN CIRCUMSTANCES, CAUSE CRACKS IN THE PLASTIC LEGS' BASE AND DAMAGE TO THE BASE ELECTRIC ELEVATOR MECHANISM. CONCLUSION: WE HAVE AN OPEN CAPA TO ADDRESS THIS ISSUE. THE CORRECTIVE ACTION INFORMING THE USERS OF THE MAXIMUM WEIGHT FOR THE COSYCOT INFANT WARMER AND TO INSPECT THE BASES OF THEIR COSYCOTS IS EXPECTED TO BE COMPLETED BY END OF 2008. THIS CORRECTIVE ACTION HAS BEEN NOTIFIED TO THE DISTRICT OFFICE.

Description of Event or Problem · 1

A HOSPITAL REPORTED THAT BASES OF THREE IW934JEU COSYCOT INFANT WARMERS HAD CRACKS. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COSYCOT INFANT WARMER FMT FISHER & PAYKEL HEALTHCARE, LTD. IW934JEU

Patients

Seq Age Sex Outcome Treatment
1