FDA Adverse Event
Malfunction
Summary report: N
GE OEC 6800
MDR report key: 1040749
·
Received May 9, 2008
Report
- Report Number
- 1720753-2008-21404
- Event Type
- Malfunction
- Date Received
- May 9, 2008
- Date of Event
- April 28, 2008
- Report Date
- May 8, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REP REPLACED THE CCD CAMERA. THE IMAGE ROTATOR NOW PERFORMS AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
CUSTOMER REPORTED THE SYSTEM DISPLAYED AN IMAGE ROTATION ERROR. THE UNIT HAD TO BE REBOOTED TO RESOLVE THE ERROR. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | 6800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |