FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1040742 · Received May 9, 2008

Report

Report Number
1720753-2008-21360
Event Type
Malfunction
Date Received
May 9, 2008
Date of Event
April 29, 2008
Report Date
May 7, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE SERVICE REP REMOVED AND REPLACED X-RAY TUBE, RAN A FILAMENT CAL WITHOUT ERROR, AND VERIFIED FLUORO AT LOW AND HIGH TECHNIQUE WITHOUT ERROR. VERIFIED SYSTEM IS OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE X-RAY TUBE IS ARCING. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1