FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1040733 · Received May 9, 2008

Report

Report Number
1720753-2008-21374
Event Type
Malfunction
Date Received
May 9, 2008
Date of Event
April 24, 2008
Report Date
May 8, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE COLLIMATOR SHUTTER MOTOR WAS FOUND TO NOT WORK. THE CUSTOMER WILL REPLACE THE COLLIMATOR ASSEMBLY. A FOLLOW UP REPORT WILL BE FILED WHEN ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9800 SYSTEM HAD BRIGHT IMAGES DURING A CASE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1