FDA Adverse Event
Malfunction
Summary report: N
ASCENSIA BREEZE2 TEST STRIPS (50)
MDR report key: 1040680
·
Received May 8, 2008
Report
- Report Number
- 1826988-2008-00467
- Event Type
- Malfunction
- Date Received
- May 8, 2008
- Date of Event
- April 10, 2008
- Report Date
- April 10, 2008
- Manufacturer
- BAYER HEALTHCARE, LLC
- Product Code
- NBW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER STATED THAT HE TESTED HIS BLOOD GLUCOSE USING HIS BREEZE2 METER AND RECEIVED A READING OF 329 MG/DL. HE RETESTED USING ANOTHER METER AND RECEIVED A READING OF 159 MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE PARKES ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASCENSIA BREEZE2 TEST STRIPS (50) | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE, LLC | 1470 | 1A5199AA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |