PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED
Report
- Report Number
- 1423500-2008-00328
- Event Type
- Malfunction
- Date Received
- May 9, 2008
- Date of Event
- April 11, 2008
- Report Date
- April 11, 2008
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- FKX
- PMA / PMN Number
- K012988
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
BAXTER REQUESTED THE DEVICE FOR EVAL; HOWEVER, THE CUSTOMER DECLINED TO MAKE THE DEVICE AVAILABLE.
THE NURSE (RN) OF A HOME PATIENT (HP) CONTACTED A TECHNICAL SERVICE REPRESENTATIVE (TSR) AND REPORTED THE HP FELT ABDOMINAL BLOATING AND DISCOMFORT, AS IF HE WERE OVERFILLED WITH FLUID, USING THE HOMECHOICE PRO AUTOMATED PERITONEAL DIALYSIS (APD) SYSTEM. THE INCIDENT WAS REVIEWED OVER THE PHONE WITH THE TSR, WHICH REVEALED THE HP DRAINED OUT 1800 MLS DURING THE APD THERAPY INITIAL DRAIN, WHICH WAS 800 MLS GREATER THAN THE CYCLER'S PROGRAMMED LAST FILL VOLUME OF 1000 MLS. ACCORDING TO THE RN, THE APD CYCLER WAS PROGRAMMED FOR 6 CYCLES OF 3000 MLS WITH A LAST FILL VOLUME OF 1000 MLS. THE RN SUSPECTED THE ADDITIONAL 800 MLS OF FLUID DRAINED WAS EITHER ACTUAL ULTRAFILTRATION REMOVED FROM THE HP'S BODY FROM NORMAL PERITONEAL DIALYSIS THERAPY OR AN INCOMPLETE DRAIN PRIOR TO THE RECEIPT OF THE LAST FILL. CYCLER DRAIN LOGIC, MOVE-ON CRITERIA FROM DRAIN TO FILL AND THE LAST MANUAL DRAIN FEATURE WAS REVIEWED WITH THE RN. THE RN HAS NOT MADE ANY THERAPY CHANGES AS A RESULT OF THIS INCIDENT, SINCE THE HP HAS NOT SHOWN UP TO CLINIC RECENTLY. THE RN STATES THAT IF THE HP CONTINUES TO EXPERIENCE ABDOMINAL BLOATING AND/OR DISCOMFORT, SHE MAY INCREASE THE CYCLER'S MINIMUM DRAIN VOLUME FROM 85% TO 100% AND IMPLEMENT A LAST MANUAL DRAIN. THERE WAS NO PT INJURY OR MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED | 78FKX | FKX | BAXTER HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | EXTRANEAL| 2.5% LO/CAL 6L DIANEAL| 1.5% LO/CAL 6L DIANEAL |