REF BALL JNT SCREWDRIVER SHAFT
Report
- Report Number
- 1020279-2020-03913
- Event Type
- Malfunction
- Date Received
- August 13, 2020
- Date of Event
- July 22, 2020
- Report Date
- August 19, 2020
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- HXX
- UDI-DI
- 03596010454201
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
RESULTS OF INVESTIGATION: THE DEVICE, INTENDED USE IN TREATMENT, WAS RETURNED FOR EVALUATION. THE VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMS THE SCREWDRIVER PART OF THE DEVICE BROKE OFF. THE BROKEN PIECE WAS NOT RETURNED WITH THE DEVICE. ALSO THE DEVICE IS ROUNDED OFF ON ONE END AND THERE ARE SEVERAL NICKS ON THE METAL OF THE DEVICE. THIS INSTRUMENT EXHIBITS SIGNS OF SIGNIFICANT USE AND WEAR. THE DEVICE MANUFACTURED IN 2013. A REVIEW OF COMPLAINT HISTORY ON THE LISTED PART REVEALED NO PRIOR COMPLAINTS FOR THE LISTED BATCH WITH THE SAME FAILURE MODE. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. THIS DEVICE IS A REUSABLE INSTRUMENTS THAT CAN BE EXPOSED TO NUMEROUS SURGERIES; DAMAGE FROM REPEATED USE CAN OCCUR. EXPECTED WEAR/TEAR IS LIKELY THE PROBABLE CAUSE OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. AT THIS TIME, WE HAVE NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINTS WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THESE COMPLAINTS; HOWEVER WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY.
IT WAS REPORTED THAT DURING AN OFFICE INSPECTION THE REF BALL JNT SCREWDRIVER SHAFT WAS FOUND TO BE BROKEN. NO CASE RELATED THEREFORE NO PATIENT INVOLVED.
IT WAS REPORTED THAT DURING AN OFFICE INSPECTION THE REF BALL JNT SCREWDRIVER SHAFT WAS FOUND TO BE BROKEN. NO CASE RELATED THEREFORE NO PATIENT INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 867850 | REF BALL JNT SCREWDRIVER SHAFT | SCREWDRIVER | HXX | SMITH & NEPHEW, INC. | 71362295 | 13DSY0002 | 03596010454201 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |