43 CM (17") APPX 2.8 ML, AMBER TRANSFER SET W/CHEMOCLAVE ADDITIVE PORT, ROTATIN
Report
- Report Number
- 9617594-2020-00317
- Event Type
- Malfunction
- Date Received
- August 13, 2020
- Date of Event
- August 5, 2020
- Report Date
- August 6, 2020
- Manufacturer
- ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
- Product Code
- ONB
- UDI-DI
- 00840619095550
- PMA / PMN Number
- K173477
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHARMACIST
Narratives
H10: NO PRODUCT SAMPLES, VIDEOS, OR PHOTOGRAPHS WERE RETURNED FOR INVESTIGATION. HOWEVER, A COMPLAINT HISTORY REVIEW SHOWED THE REPORTED ISSUE OF LEAKAGE AT THE CONNECTION WAS PREVIOUSLY CONFIRMED FOR THE SAME MALE LUER PURCHASED PART LOT NUMBER. THE REPORTED COMPLAINT OF LEAKAGE CAN BE CONFIRMED. THE PROBABLE CAUSE OF THE LEAKAGE IS DUE TO A MOLDING ANOMALY ON THE MALE LUER POST CREATED DURING THE MOLDING PROCESS AT THE SUPPLIERS FACILITY. THE DEVICE HISTORY REVIEW FOR LOT NUMBER 4734811 WAS REVIEWED AND THERE WERE NO RELEVANT NON-CONFORMANCES FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.
THE SAMPLE WAS DISCARDED. WITHOUT THE RETURN OF THE SAMPLE A COMPREHENSIVE FAILURE INVESTIGATION CANNOT BE PERFORMED AND A CAUSE CANNOT BE DETERMINED. CONCOMITANT PRODUCTS - 28 CM (11") ADD-ON SET W/2 CLAVE®, VENTED CAP, LN 011-H3314, LOT NUMBER 4866115, MFR ICU MEDICAL; PRIMARY PLUM SET, 15 MICRON FILTER IN SIGHT CHAMBER, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 272 CM, LN 140019291, LOT NUMBER 4495817, MFR ICU MEDICAL; PLUM 360 PUMP, SN (B)(4), LN 30010-29-13, MFR ICU MEDICAL, OXALIPLATINE/FLUOROURACILE, MFR UNK.
THE EVENT INVOLVED A 43 CM (17") AMBER TRANSFER SET W/CHEMOCLAVE® ADDITIVE PORT, ROTATING LUER, FILTER CAP, 10 UNITS WHERE THERE WAS A LEAK NOTED DURING A CHEMOTHERAPY INFUSION. THE DEVICE WAS CONNECTED TO AN ADD-ON SET AND A PRIMARY PLUMSET AND WAS INFUSING AN UNSPECIFIED CONCENTRATION OF OXALIPLATINE/FLUOROURACILE VIA A PLUM 360 PUMP. A MAJOR LEAK OF CHEMOTHERAPY WAS FOUND TO HAVE LEAKED ON THE FLOOR FROM THE CONNECTION BETWEEN THE TRANSFER SET AND ADD-ON SET AND WAS CLEANED PER PROTOCOL. THERE WAS NO MEDICAL INTERVENTION REQUIRED. ALTHOUGH THERE WAS A DELAY IN THERAPY REPORTED AND AN INCOMPLETE ADMINISTRATION OF THE MEDICATION, THERE WAS NO PATIENT HARM AND NO BLOOD LOSS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 867699 | 43 CM (17") APPX 2.8 ML, AMBER TRANSFER SET W/CHEMOCLAVE ADDITIVE PORT, ROTATIN | CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM | ONB | ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. | 4734811 | 00840619095550 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SEE H10 |