FDA Adverse Event Malfunction Summary report: N

43 CM (17") APPX 2.8 ML, AMBER TRANSFER SET W/CHEMOCLAVE ADDITIVE PORT, ROTATIN

MDR report key: 10404731 · Received August 13, 2020

Report

Report Number
9617594-2020-00317
Event Type
Malfunction
Date Received
August 13, 2020
Date of Event
August 5, 2020
Report Date
August 6, 2020
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
ONB
UDI-DI
00840619095550
PMA / PMN Number
K173477
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 0

H10: NO PRODUCT SAMPLES, VIDEOS, OR PHOTOGRAPHS WERE RETURNED FOR INVESTIGATION. HOWEVER, A COMPLAINT HISTORY REVIEW SHOWED THE REPORTED ISSUE OF LEAKAGE AT THE CONNECTION WAS PREVIOUSLY CONFIRMED FOR THE SAME MALE LUER PURCHASED PART LOT NUMBER. THE REPORTED COMPLAINT OF LEAKAGE CAN BE CONFIRMED. THE PROBABLE CAUSE OF THE LEAKAGE IS DUE TO A MOLDING ANOMALY ON THE MALE LUER POST CREATED DURING THE MOLDING PROCESS AT THE SUPPLIERS FACILITY. THE DEVICE HISTORY REVIEW FOR LOT NUMBER 4734811 WAS REVIEWED AND THERE WERE NO RELEVANT NON-CONFORMANCES FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.

Additional Manufacturer Narrative · 1

THE SAMPLE WAS DISCARDED. WITHOUT THE RETURN OF THE SAMPLE A COMPREHENSIVE FAILURE INVESTIGATION CANNOT BE PERFORMED AND A CAUSE CANNOT BE DETERMINED. CONCOMITANT PRODUCTS - 28 CM (11") ADD-ON SET W/2 CLAVE®, VENTED CAP, LN 011-H3314, LOT NUMBER 4866115, MFR ICU MEDICAL; PRIMARY PLUM SET, 15 MICRON FILTER IN SIGHT CHAMBER, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 272 CM, LN 140019291, LOT NUMBER 4495817, MFR ICU MEDICAL; PLUM 360 PUMP, SN (B)(4), LN 30010-29-13, MFR ICU MEDICAL, OXALIPLATINE/FLUOROURACILE, MFR UNK.

Description of Event or Problem · 1

THE EVENT INVOLVED A 43 CM (17") AMBER TRANSFER SET W/CHEMOCLAVE® ADDITIVE PORT, ROTATING LUER, FILTER CAP, 10 UNITS WHERE THERE WAS A LEAK NOTED DURING A CHEMOTHERAPY INFUSION. THE DEVICE WAS CONNECTED TO AN ADD-ON SET AND A PRIMARY PLUMSET AND WAS INFUSING AN UNSPECIFIED CONCENTRATION OF OXALIPLATINE/FLUOROURACILE VIA A PLUM 360 PUMP. A MAJOR LEAK OF CHEMOTHERAPY WAS FOUND TO HAVE LEAKED ON THE FLOOR FROM THE CONNECTION BETWEEN THE TRANSFER SET AND ADD-ON SET AND WAS CLEANED PER PROTOCOL. THERE WAS NO MEDICAL INTERVENTION REQUIRED. ALTHOUGH THERE WAS A DELAY IN THERAPY REPORTED AND AN INCOMPLETE ADMINISTRATION OF THE MEDICATION, THERE WAS NO PATIENT HARM AND NO BLOOD LOSS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
867699 43 CM (17") APPX 2.8 ML, AMBER TRANSFER SET W/CHEMOCLAVE ADDITIVE PORT, ROTATIN CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM ONB ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 4734811 00840619095550

Patients

Seq Age Sex Outcome Treatment
1 SEE H10