ATN LAG SCREW 110MM STERILE
Report
- Report Number
- 1818910-2008-01341
- Event Type
- Injury
- Date Received
- May 8, 2008
- Date of Event
- April 16, 2008
- Report Date
- April 23, 2008
- Manufacturer
- DEPUY ACE S.A. R.L.
- Product Code
- HSB
- PMA / PMN Number
- K042325
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
EXAMINATION WAS NOT POSSIBLE, AS THE PRODUCT WAS NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE FOUND NO PRIOR REPORTS FOR BOTH REPORTED PART AND LOT NUMBER COMBINATIONS. REVIEW OF THE DEVICE HISTORY RECORDS BY LELOCLE FOUND NO DISCREPANCIES. ALTHOUGH THE COMPLAINT IS NON-VERIFIABLE, INITIAL IMPLANT PLACEMENT AND BONE QUALITY ARE CONTRIBUTING FACTORS. ALSO, THE SURGICAL TECHNIQUE FOR THE ATN SYSTEM RECOMMENDS TAPPING OF THE FEMORAL HEAD BEFORE INSERTION OF THE LAG SCREW. THE ATN LAG SCREW IS NOT SELF-TAPPING. IF THE FEMORAL HEAD IS NOT TAPPED BEFORE LAG SCREW INSERTION, THE LAG SCREW FORCES THE BONE IN THE FEMORAL HEAD TO BE PUSHED AWAY FROM THE LAG SCREW THREAD. THE THREADS WILL NOT PURCHASE INTO THE BONE AND THERE IS NOT ADEQUATE SUPPORT TO HOLD THE LAG SCREW IN THE FEMORAL HEAD. IF THE LAG SCREW IS INADEQUATELY FIXED INTO THE FEMORAL HEAD LAG SCREW BACKOUT CAN OCCUR. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD THE PRODUCT AND/OR ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.
IT WAS FOUND THAT THE LAG SCREW WAS BACKED OUT TWO MONTHS AND A HALF AFTER THE SURGERY. THE PT HAD A PAIN. ANOTHER SURGERY WAS DONE TO REMOVE THE LAG SCREW OUT OF THE PATIENT'S BODY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATN LAG SCREW 110MM STERILE | 87HSB | HSB | DEPUY ACE S.A. R.L. | NA | DFDCJM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Required Intervention |