FDA Adverse Event Malfunction Summary report: N

PULSE GEN MODEL 101

MDR report key: 1040372 · Received May 6, 2008

Report

Report Number
1644487-2008-01089
Event Type
Malfunction
Date Received
May 6, 2008
Date of Event
January 15, 2008
Report Date
April 30, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

IT WAS REPORTED TO MANUFACTURER THAT THE VNS PT HAD THEIR GENERATOR REPLACED DUE TO END OF SERVICE. A BATTERY LIFE CALCULATION WAS PERFORMED AND WAS CALCULATED TO BE -2.60 YEARS UNTIL ERI (THE IMPLANT TIME EXCEED THE BATTERY LIFE ESTIMATE). THE GENERATOR WAS RETURNED TO MANUFACTURER FOR ANALYSIS. THE ANALYSIS FINDINGS SUPPORTED THE REPORTED END OF SERVICE CLAIM BASED ON BATTERY VOLTAGE MEASURING 1.878 VOLTS (DEPLETED). DURING TESTING, TRANSISTOR Q10 WAS FOUND TO EXHIBIT AN OUT-OF-LIMIT (HIGHER) LEAKAGE CURRENT AT TEST CHAMBER TEMPERATURE. THIS LEAKAGE INCREASED THE MODULE'S SUPPLY CURRENT 1MA CONSUMPTION BY APPROXIMATELY 5.137UA OVER THE HIGH LIMIT (SPEC 23.000 - 38.000MA) AND COULD POTENTIALLY BE A CONTRIBUTING FACTOR TO THE END OF SERVICE OF THE GENERATOR. HOWEVER, RESULTS OF THE BATTERY LONGEVITY PREDICTION CONFIRM THAT THE END OF SERVICE CONDITION WAS AN EXPECTED EVENT. WHEN THE Q10 COMPONENT WAS REPLACED, THE DEVICE WAS TESTED AGAIN, AND SUBSEQUENTLY MET MANUFACTURERS TEST SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PULSE GEN MODEL 101 LYJ CYBERONICS, INC. 101 1726