FDA Adverse Event Injury Summary report: N

ICM (IMPLANTABLE COLLAMER LENS)

MDR report key: 1040288 · Received May 8, 2008

Report

Report Number
2023826-2008-00645
Event Type
Injury
Date Received
May 8, 2008
Date of Event
January 21, 2008
Report Date
April 14, 2008
Manufacturer
STAAR SURGICAL
Product Code
MTA
PMA / PMN Number
P030016
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: CONCLUSIONS: AT THE CONCLUSION OF THE ORIGINAL INVESTIGATION OPENED FOR THE ISSUE OF ICL VAULTING (BOTH INADEQUATE AND EXCESSIVE), IT WAS DETERMINED THAT THE MOST LIKELY ROOT CAUSE FOR BOTH INADEQUATE AND EXCESSIVE VAULT IS DIFFICULTY IN CLINICAL SIZING. IN NO CASE HAS A RETURNED LENS EXHIBITED A LENGTH MEASUREMENT OUTSIDE THE EXPECTED RANGE ACCORDING TO THE LABELED LENGTH. PRE-OPERATIVE MEASUREMENT TECHNOLOGY AVAILABLE TO CLINICIANS IN THE PAST DID NOT PROVIDE A DIRECT MEASUREMENT OF THE SULCUS, TO WHICH THE ICL SHOULD BE SIZED. THE CURRENT PRACTICE IN SELECTING AN APPROPRIATE ICL FOR A PT IS TO MEASURE WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH AND, THROUGH CORRELATIVE ALGORITHM PREDICT THE LENGTH OF THE ICL THAT WILL BRIDGE THE SULCUS. THIS METHOD OF SIZING BASED UPON WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH MEASUREMENTS WAS USED IN THE FDA CLINICAL TRIAL AND IS RECOMMENDED IN THE CURRENT LABELING OF THE ICL. IF THE PREDICTED LENGTH IS TOO SHORT, THE RESULT MAY BE AN INADEQUATE VAULT. IF PREDICTED LENGTH IS TOO LONG, THE RESULT MAY BE AN EXCESSIVE VAULT.

Description of Event or Problem · 1

THE REPORTER STATED, THAT THE SURGEON INSERTED AN (IMPLANTABLE COLLAMER LENS) MODEL ICM 120V4 IN 2008 AND DUE TO AN EXCESSIVE VAULT OF THE LENS CAUSING A NARROWING OF THE ANGLE AND SHALLOWING OF THE ANTERIOR CHAMBER DEPTH, THE SURGEON EXPLANTED THE LENS FOURTEEN DAYS LATER AND EXCHANGED IT FOR A SHORTER LENGTH LENS. TWO IRIDOTOMIES WERE PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ICM (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS MTA STAAR SURGICAL ICM120V4 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention CARTRIDGE MODEL: SFC-45FP| INJECTOR MODEL: MSI-PF| FOAM TIP PLUNGER: FTP