FDA Adverse Event Injury Summary report: N

PERF ANAT TIBIAL STEM 120X14X5

MDR report key: 10402742 · Received August 13, 2020

Report

Report Number
0009610576-2020-00009
Event Type
Injury
Date Received
August 13, 2020
Date of Event
June 8, 2020
Report Date
January 29, 2021
Manufacturer
BIOMET SPAIN, S.L.
Product Code
KRR
PMA / PMN Number
N/A
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

(B)(4). D11: ASSOCIATED MEDICAL PRODUCTS REF. 178892; LOT 0000151670; DESCRIPTION PS CONSTRAINED INSERT LGE 22MM W/SCREW. REF. 177284; LOT 2007050790; DESCRIPTION MODULAR TIBIAL TRAY LARGE 76 MM CEMENTED. ADDITIONAL INFORMATION REGARDING THE INVOLVED LOTS DETECTED DURING THE DOCUMENTAL INVESTIGATION.

Description of Event or Problem · 0

REVISION DUE TO INSTABILITY, LOOSENING OF FEMALE SCREW, OSTEOLYSIS, POLYESILE METAL DISPOSAL.

Additional Manufacturer Narrative · 1

(B)(4). ASSOCIATED MEDICAL PRODUCTS: REF. 177284; LOT 2007060790; DESCRIPTION MODULAR TIBIAL TRAY LGE 76 MM CEMENTED. REF. 178892; LOT 151670; DESCRIPTION PS CONSTRAINED INSERT LGE 22MM W/SCREW. REF. 7806-16-150; LOT 2007030389; DESCRIPTION FEMORAL STEM 16X150MM W/SCREW. REF. 177043; LOT 05037715; DESCRIPTION PS FEMORAL COMPONENT CLOSED BOX MED LEFT 70 MM CEMENTED. (B)(6). FOREIGN REPORT SOURCE: (B)(6). THIS PRODUCT IS MANUFACTURED BY BIOMET (B)(4) AND IS NOT CLEARED OR DISTRIBUTED IN THE U.S. HOWEVER, THIS REPORT IS BEING SUBMITTED AS ZIMMER BIOMET WARSAW MANUFACTURES A SIMILAR DEVICE THAT IS CLEARED OR DISTRIBUTED IN THE UNITED STATES UNDER 510(K) NUMBER K143192. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0009610576 - 2020 - 00007 AND 0009610576 - 2020 - 00008.

Description of Event or Problem · 1

IT WAS REPORTED BY THE (B)(6) PATIENT UNDERWENT TOTAL KNEE ARTHROPLASTY ON AN UNKNOWN DATE IN 2008. SUBSEQUENTLY, PATIENT UNDERWENT A REVISION (B)(6) 2020 DUE TO INSTABILITY OF TOTAL REVISION KNEE PROSTHESIS IMPLEMENTED IN 2008 DUE TO THE BREAKING OF THE CENTRAL STABILIZATION POST OF THE SAME, WITH LOOSENING OF THE FEMALE STEM SCREW, PRESENCE OF MULTIPLE OSTEOLYSIS, POLYESILE METAL DISPOSAL. THE CENTRAL POST WAS FOUND IN THE HOLLOW POPLITEO. RECONSTRUCTION IS NEEDED WITH CONSTRAINED IMPLANT AND INCREASES OF TRABECULAR METAL GIVEN THE LOSS OF SUBSTANCE AND THE AFFECTION OF THE LIG. SIDES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
866449 PERF ANAT TIBIAL STEM 120X14X5 PROSTHESIS, KNEE KRR BIOMET SPAIN, S.L. N/A 2007010785

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization SEE H10 NARRATIVE