TUBE, GASTROINTESTINAL (AND ACCESS)
Report
- Report Number
- 2523003-2008-00075
- Event Type
- Injury
- Date Received
- May 8, 2008
- Date of Event
- April 2, 2008
- Report Date
- April 15, 2008
- Manufacturer
- C.R. BARD, INC. (BASD)
- Product Code
- KNT
- PMA / PMN Number
- K915837
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE COMPLAINT OF A BREAK IN THE RETENTION DOME IS CONFIRMED, USER RELATED. UPON RECEIPT, A PARTIAL SEMI-CIRCULAR BREAK IN THE RETENTION DOME WAS OBSERVED. THE DULL AND ROUNDED VENEER IDENTIFIED AT THE DOME SITE ARE TRAITS THAT ARE INDICATIVE OF EXCESSIVE FORCE THAT WAS APPLIED DURING THE REMOVAL OF THIS DEVICE. THE LACK OF ANY ASSOCIATED DAMAGE SUGGESTS THE TEAR IN THE DOME WAS CAUSED BY STRETCHING THE DOME MATERIAL BEYOND ITS ELASTIC LIMITS DURING REMOVAL. THE COMPLAINANT INDICATES THE COMPLAINT SAMPLE HAD BEEN IN PLACE FOR APPROX 195 DAYS PRIOR TO THE COMPLAINT INCIDENT. GROSS VISUAL AND MICROSCOPIC EXAMINATIONS SHOW NO EVIDENCE OF A DEFECT OR DEFORMITY RELATED TO THE MANUFACTURING PROCESS. A CHR IS NOT POSSIBLE, AS NO MANUFACTURING LOT NUMBER HAS BEEN PROVIDED BY THE COMPLAINANT.
A BREAK IN THE RETENTION DOME DURING TRACTION REMOVAL. THE INDWELLING PERIOD WAS 195 DAYS. THE RETENTION DOME DROPPED INTO THE STOMACH AND REMOVED ENDOSCOPICALLY AFTERWARDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TUBE, GASTROINTESTINAL (AND ACCESS) | KNT | C.R. BARD, INC. (BASD) | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |