FDA Adverse Event Injury Summary report: N

STARCLOSE VASCULAR CLOSURE SYSTEM

MDR report key: 1040171 · Received May 8, 2008

Report

Report Number
2953144-2008-00863
Event Type
Injury
Date Received
May 8, 2008
Date of Event
April 9, 2008
Report Date
April 15, 2008
Manufacturer
ABBOTT VASCULAR REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P050007
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED.

Description of Event or Problem · 1

DEVICE MALFUNCTION: DIFFICULT TO REMOVE. TIME OF DEVICE MALFUNCTION: DURING VESSEL CLOSURE. SYMPTOMS/AE: FAILURE TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, AFTER THE FOURTH CLICK THE DEVICE BECAME STUCK IN THE WOUND TRACT. THE DEVICE WAS REMOVED USING COUNTER-TENSION AND HEMOSTASIS WAS ACHIEVED USING MANUAL COMPRESSION. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. THROUGH REQUESTED, NO ADDITIONAL INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE VASCULAR CLOSURE SYSTEM MGB ABBOTT VASCULAR REDWOOD CITY NA 62173-6H

Patients

Seq Age Sex Outcome Treatment
1 29 YR Required Intervention