FDA Adverse Event Injury Summary report: N

DRIVE SHAFT FOR RIA 2 520MM

MDR report key: 10397683 · Received August 12, 2020

Report

Report Number
8030965-2020-05908
Event Type
Injury
Date Received
August 12, 2020
Date of Event
July 31, 2020
Report Date
July 31, 2020
Manufacturer
OBERDORF SYNTHES PRODUKTIONS GMBH
Product Code
HTO
UDI-DI
07612334143037
PMA / PMN Number
K111437
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. VISUAL INSPECTION: DRIVE SHAFT F/RIA 2 L520 (PART. NO: 03.404.035, LOT. NO: H860686, QTY: 1) WAS RETURNED AND RECEIVED AT US CUSTOMER QUALITY (CQ). UPON VISUAL INSPECTION AT CQ, IT IS OBSERVED THAT THERE ARE FEW SCRATCHES ALL OVER THE BODY WHICH WOULD NOT IMPACT THE FUNCTIONALITY OF THE DEVICE. NO OTHER PHYSICAL DAMAGE WAS OBSERVED WITH THE DEVICE. FUNCTIONAL TEST: FUNCTIONAL TESTING OF THE RECEIVED DEVICE CANNOT BE PERFORMED DUE TO THE BROKEN REAMER HEADS (MATING DEVICES). CAN THE COMPLAINT BE REPLICATED WITH THE RETURNED DEVICES? UNABLE TO PERFORM. DIMENSIONAL INSPECTION: (CALIPERS: CA802) DIMENSIONAL INSPECTION OF THE RETURNED DEVICE WAS PERFORMED AT CUSTOMER QUALITY (CQ). THE OUTER DIAMETER OF THE AUGER OF THE REAMER SHAFT DEVICE WAS INTENDED TO BE MEASURING FROM 6.90MM TO 7.10MM AND IT WAS MEASURED TO BE 6.95MM WHICH IS WITHIN SPECIFICATION AS PER THE DRAWING. DOCUMENT/SPECIFICATION REVIEW: THE FOLLOWING DRAWINGS WERE REVIEWED DURING THE INVESTIGATION: -AUGER, RIA 2: 03_404_035_07 REV. B AND REV. C, -DRIVE SHAFT, RIA 2: 03_404_035 REV. B AND REV. C. NO DESIGN ISSUES OR DISCREPANCIES WERE NOTED DURING THE INVESTIGATION. COMPLAINT CONFIRMED? NO REPORTED EMBEDDED CONDITION CANNOT BE CONFIRMED AS THE DEVICE DOESN¿T HAVE ANY BROKEN FEATURES. INVESTIGATION CONCLUSION: THE COMPLAINT COULD NOT BE CONFIRMED FOR DRIVE SHAFT F/RIA 2 L520 (PART. NO: 03.404.035, LOT. NO: H860686). A DEFINITIVE ROOT CAUSE FOR THE REPORTED PROBLEM COULD NOT BE DETERMINED. DURING THE INVESTIGATION NO UNIDENTIFIED PRODUCT DESIGN/MANUFACTURING ISSUES OR DISCREPANCIES WERE OBSERVED THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H6: INVESTIGATION SUMMARY PRODUCT WAS NOT RETURNED. REVIEWING ATTACHED PICTURE, METAL FRAGMENTS REMAINED IN PATIENTS BONE COULD BE CONFIRMED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT: PART # 03.404.035, SYNTHES LOT # H860686, SUPPLIER LOT # H860686, RELEASE TO WAREHOUSE DATE: NOV 05, 2019 , SUPPLIER: (B)(4). NO NCR'S WERE GENERATED DURING PRODUCTION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL DEVICE PRODUCT CODES: HRX. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/ INVESTIGATION, BUT HAS YET TO BE RECEIVED. REPORTER IS A SYNTHES EMPLOYEE. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, DURING AN RIA II SURGERY THE DRILL HEAD LOOSENED ITSELF FROM THE DRILLING SHAFT. THE DRILLING HEAD COULD BE RECOVERED WITH THE 2,5MM DRILL MANTEL WHICH WAS CONNECTED TO THE OLIVE. THE DRILLING HEAD WAS REATTACHED BUT CAME LOOSE AGAIN AND WAS RECOVERED AGAIN. A NEW DRILL HEAD WITH A DIAMETER OF 12MM WAS USED AFTER THAT WHICH ALSO LOOSENED ITSELF FROM THE DRILLING SHAFT. ON THE X-RAY IT APPEARED THAT METAL FRAGMENTS HAD REMAINED AT THE ENTRANCE OF THE MEDULLARY CANAL. THERE WAS A SURGICAL DELAY OF FIFTEEN (15) MINUTES. ENOUGH MATERIAL COULD BE HARVESTED FROM THE MEDULLARY CAVITY TO COMPLETE THE SURGERY SUCCESSFULLY. ANOTHER PIECE DETACHED FROM THE DRILLING SHAFT END WHERE THE DRILL HEAD IS LOCKED IN PLACE. THE PATIENT WAS REPORTED AS GOOD AFTER THE PROCEDURE. CONCOMITANT DEVICES REPORTED: REAMER HEAD F/RIA 2 Ø12.5 (PART NUMBER 03.404.021S, LOT 26P2108, QUANTITY 1), REAMER HEAD F/RIA 2 Ø12 (PART NUMBER 03.404.020S, LOT 39P1189, QUANTITY 1). THIS REPORT INVOLVES ONE (1) DRIVE SHAFT FOR RIA 2 520MM. THIS IS REPORT 3 OF 3 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
863373 DRIVE SHAFT FOR RIA 2 520MM REAMER HTO OBERDORF SYNTHES PRODUKTIONS GMBH H860686 07612334143037

Patients

Seq Age Sex Outcome Treatment
1 48 YR Required Intervention REAMER HEAD F/RIA 2 Ø12| REAMER HEAD F/RIA 2 Ø12.5