TRUE METRIX
Report
- Report Number
- 1000113657-2020-00564
- Event Type
- Malfunction
- Date Received
- August 12, 2020
- Date of Event
- July 15, 2020
- Report Date
- September 18, 2020
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
SECTIONS WITH ADDITIONAL INFORMATION AS OF (B)(6) 2020: H6: UPDATED FDA'S METHOD, RESULT, AND CONCLUSION CODES H10: METER WAS RETURNED FOR EVALUATION; NO DEFECT WAS DETECTED. TEST STRIPS WERE NOT RETURNED FOR EVALUATION.
INTERNAL REPORT REFERENCE NUMBER: (B)(4). PRODUCTS WERE RETURNED AND PENDING QC EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 27-JUL-2020 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER AND STATED REPLACEMENT PRODUCT RESOLVED INITIAL CONCERN.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. PHARMACIST IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 257, 172, 162 AND 164 MG/DL. THE CUSTOMERS EXPECTED FASTING BLOOD GLUCOSE TEST RESULT IS 126 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURERS EXPIRATION DATE IS 11/20/2021 AND OPEN VIAL DATE IS 07/07/2020. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 862732 | TRUE METRIX | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TMX HEB 50CTMG/DL | MX4246S | 00021292007836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |