FDA Adverse Event Injury Summary report: N

ERBECRYO 2

MDR report key: 10395220 · Received August 12, 2020

Report

Report Number
9610614-2020-00025
Event Type
Injury
Date Received
August 12, 2020
Date of Event
July 27, 2020
Report Date
August 12, 2020
Manufacturer
ERBE ELEKTROMEDIZIN GMBH
Product Code
GEH
PMA / PMN Number
K190651
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CRYOSURGICAL UNIT WAS THOROUGHLY INSPECTED/TESTED [NOTE: THE SINGLE USE CRYOPROBE WAS DISCARDED AND NOT AVAILABLE FOR AN EVALUATION. IT WAS REPORTED THOUGH THAT THE PROBE WORKED WELL AND FORMED AN ICE BALL AS INTENDED WITHOUT LEAKING.]. A TECHNICAL SAFETY CHECK WAS PERFORMED ON THE UNIT. THIS INCLUDED AN ELECTRICAL SAFETY CHECK, A FUNCTIONAL CHECK OF EACH OF THE EQUIPMENT'S FEATURES, AND A POWER OUTPUT CHECK. ALL FEATURES WERE/ARE FUNCTIONING PROPERLY WITHIN SPECIFICATIONS FOR THE DEVICE. NO ANOMALIES WERE FOUND IN THE REVIEW OF THE CRYOSURGICAL UNIT'S DEVICE HISTORY RECORD (DHR) [NOTE: THE DHR FOR THE CRYOPROBE DEMONSTRATED THAT THE PROBE PASSED LEAK TESTING.]. IN CONCLUSION, NO EQUIPMENT PROBLEM WAS FOUND THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE EVENT. BASED UPON THE DESCRIPTION OF THE EVENT AND A REVIEW OF LITERATURE, INJURIES TO LUNG PARENCHYMA CAN LEAD TO A GAS EMBOLISM WHICH COULD CAUSE CARDIAC ARREST. NO TRENDS HAVE BEEN IDENTIFIED WITH THIS INCIDENT. ERBE USA, INC. IS NOW CLOSING THE FILE ON THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT INCIDENT OCCURRED WITH THE CRYOSURGICAL UNIT DURING A BIOPSY IN THE TRACHEOBRONCHIAL TREE. THE UNIT WAS USED WITH A SINGLE USE CRYOPROBE (PART NUMBER 20402-401, LOT NUMBER WO315706). DURING THE PROCEDURE, AN EMBOLISM OCCURRED AND THE PATIENT WENT INTO CARDIAC ARREST. THROUGH EMERGENCY TREATMENT, THE PATIENT WAS REVIVED. NOTE: THE UNIT AND CRYOPROBE WERE DISTRIBUTED TO A MEDICAL CENTER IN (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
861048 ERBECRYO 2 CRYOSURGICAL UNIT AND ACCESSORIES GEH ERBE ELEKTROMEDIZIN GMBH ERBECRYO 2

Patients

Seq Age Sex Outcome Treatment
1 50 YR Life Threatening| R