ERBECRYO 2
Report
- Report Number
- 9610614-2020-00025
- Event Type
- Injury
- Date Received
- August 12, 2020
- Date of Event
- July 27, 2020
- Report Date
- August 12, 2020
- Manufacturer
- ERBE ELEKTROMEDIZIN GMBH
- Product Code
- GEH
- PMA / PMN Number
- K190651
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE CRYOSURGICAL UNIT WAS THOROUGHLY INSPECTED/TESTED [NOTE: THE SINGLE USE CRYOPROBE WAS DISCARDED AND NOT AVAILABLE FOR AN EVALUATION. IT WAS REPORTED THOUGH THAT THE PROBE WORKED WELL AND FORMED AN ICE BALL AS INTENDED WITHOUT LEAKING.]. A TECHNICAL SAFETY CHECK WAS PERFORMED ON THE UNIT. THIS INCLUDED AN ELECTRICAL SAFETY CHECK, A FUNCTIONAL CHECK OF EACH OF THE EQUIPMENT'S FEATURES, AND A POWER OUTPUT CHECK. ALL FEATURES WERE/ARE FUNCTIONING PROPERLY WITHIN SPECIFICATIONS FOR THE DEVICE. NO ANOMALIES WERE FOUND IN THE REVIEW OF THE CRYOSURGICAL UNIT'S DEVICE HISTORY RECORD (DHR) [NOTE: THE DHR FOR THE CRYOPROBE DEMONSTRATED THAT THE PROBE PASSED LEAK TESTING.]. IN CONCLUSION, NO EQUIPMENT PROBLEM WAS FOUND THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE EVENT. BASED UPON THE DESCRIPTION OF THE EVENT AND A REVIEW OF LITERATURE, INJURIES TO LUNG PARENCHYMA CAN LEAD TO A GAS EMBOLISM WHICH COULD CAUSE CARDIAC ARREST. NO TRENDS HAVE BEEN IDENTIFIED WITH THIS INCIDENT. ERBE USA, INC. IS NOW CLOSING THE FILE ON THIS EVENT.
IT WAS REPORTED THAT A PATIENT INCIDENT OCCURRED WITH THE CRYOSURGICAL UNIT DURING A BIOPSY IN THE TRACHEOBRONCHIAL TREE. THE UNIT WAS USED WITH A SINGLE USE CRYOPROBE (PART NUMBER 20402-401, LOT NUMBER WO315706). DURING THE PROCEDURE, AN EMBOLISM OCCURRED AND THE PATIENT WENT INTO CARDIAC ARREST. THROUGH EMERGENCY TREATMENT, THE PATIENT WAS REVIVED. NOTE: THE UNIT AND CRYOPROBE WERE DISTRIBUTED TO A MEDICAL CENTER IN (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 861048 | ERBECRYO 2 | CRYOSURGICAL UNIT AND ACCESSORIES | GEH | ERBE ELEKTROMEDIZIN GMBH | ERBECRYO 2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Life Threatening| R |