FDA Adverse Event Malfunction Summary report: N

CONNECTOR BRACKETS RIGHT

MDR report key: 10390155 · Received August 11, 2020

Report

Report Number
3007802293-2020-00014
Event Type
Malfunction
Date Received
August 11, 2020
Date of Event
July 16, 2020
Report Date
October 15, 2020
Manufacturer
HANDICARE USA INC.
Product Code
FSA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PRELIMINARY PHOTOS AND A DESCRIPTION WERE SENT TO THE PROJECT MANAGEMENT GROUP, WHO IDENTIFIED SEVERAL ISSUES WITH THE INSTALLATION. HANDICARE VISITED THE HOME AND CONFIRMED SEVERAL ISSUES. THE INSTALLATION ISSUES IDENTIFIED, MOST SIGNIFICANTLY THE LONG RODS WITHOUT LATERAL BRACING, ALLOWED THE RAIL PIECES TO SEPARATE ENOUGH FOR THE UNIT TO TRAVEL OUT OF THE TRACK. THESE ISSUES WOULD BE NOTICEABLE TO THE USER. THE ROOT CAUSE OF THIS INCIDENT IS INSTALLATION ERROR. THE CEILING LIFT WAS REPLACED WITH A NEW UNIT. ON (B)(6) 2020, HANDICARE REPLACED SEVERAL COMPONENTS TO STABILIZE THE SYSTEM, UNTIL THE FULL SYSTEM COULD BE REPLACED COMPLETED (B)(6) 2020. THE COMPANY RESPONSIBLE FOR THE ORIGINAL INSTALLATION OF THE SYSTEM IS NO LONGER IN BUSINESS. THE INSTALLATION MANUAL IS WELL DESCRIBED. NO FURTHER ACTION WILL BE TAKEN.

Description of Event or Problem · 1

A C-450 UNIT FELL OUT OF THE CEILING TRACK SYSTEM WHILE IN USE. THE USER FELL TO THE FLOOR AND HIT HIS HEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
855052 CONNECTOR BRACKETS RIGHT CEILING TRACK ACCESSORY FSA HANDICARE USA INC. 360465

Patients

Seq Age Sex Outcome Treatment
1 32 YR