FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY

MDR report key: 10384508 · Received August 10, 2020

Report

Report Number
3013756811-2020-82816
Event Type
Malfunction
Date Received
August 10, 2020
Date of Event
July 18, 2020
Report Date
August 10, 2020
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613205
PMA / PMN Number
DEN180058
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN OCCLUSION ALARM OCCURRED. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 200-379 MG/DL. DURING TROUBLESHOOTING, IT WAS DETERMINED THAT THE CARTRIDGE WAS THE CAUSE OF THE OCCLUSION. REPORTEDLY, THE CARTRIDGE WAS CHANGED TO ADDRESS THE ISSUE AND INSULIN DELIVERY WAS RESUMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
848182 T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 M542970 00850006613205

Patients

Seq Age Sex Outcome Treatment
1 57 YR INSULIN: NOVOLOG / NOVORAPIDINFUSION SET: AUTOSOF