FDA Adverse Event Malfunction Summary report: N

COSYCOT INFANT WARMER

MDR report key: 1038425 · Received May 5, 2008

Report

Report Number
9611451-2008-00269
Event Type
Malfunction
Date Received
May 5, 2008
Date of Event
September 9, 2007
Report Date
September 10, 2007
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
FMT
PMA / PMN Number
K971695
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL MANUFACTURING DATES: 09/14/2005, 09/14/2005. AN INVESTIGATION WAS CARRIED OUT BASED ON THE EVENT DESCRIPTION, PHOTOS AND RESULTS OF PREVIOUS INVESTIGATION ON SIMILAR COMPLAINTS, AS THE COMPLAINT DEVICE HAD NOT BEEN RETURNED FOR EVALUATION. CRACKING OF THE INFANT WARMER HEAD IS USUALLY DUE TO THE USE OF INCOMPATIBLE CLEANING SOLUTIONS AND DOING THE CLEANING BEFORE THE WARMER HEAD HAS BEEN COOLED. IT WAS REPORTED THAT THE HOSPITAL WAS COOLING THE WARMER HEADS BEFORE DOING THE CLEANING. FOR CLEANING AND DISINFECTING, THEY WERE USING INCIDIN FOAM, WHICH IS AN ALCOHOLIC DISINFECTANT CONTAINING GLUCOPROTAMIN. THE LATTER IS AN ANTIMICROBIAL INGREDIENT, WHICH IS REPORTED TO HAVE CONTAINED AMINE. THE OPERATING MANUAL STATES THAT CLEANING OF ALL PARTS OF THE INFANT WARMER AND ACCESSORIES SHOULD BE DONE AT NORMAL ROOM TEMPERATURE, AROUND 23 DEGREES CELSIUS, AND TO AVOID CLEANING SOLUTIONS CONTAINING AROMATIC HYDROCARBONS, AMMONIA, AMINES OR AQUEOUS SOLUTIONS AS THESE MAY CAUSE CHEMICAL REACTIONS TO THE POLYCARBONATE PLASTIC MATERIAL. ALL SURFACES OF THE WARMER HEAD MAY BE CLEANED WITH DETERGENT BASED SOLUTION WITH MAXIMUM CONCENTRATION OF 2% IN WATER. CONCLUSION: CRACKING OF THE INFANT WARMER HEAD, DUE TO THE USE OF NON-COMPATIBLE CLEANING MATERIALS, IS A KNOWN PROBLEM. WE HAVE AN OPEN CAPA TO EXAMINE FURTHER COMPATIBLE CLEANING SOLUTIONS.

Description of Event or Problem · 1

A HOSPITAL REPORTED THAT THREE IW931 COSYCOT INFANT WARMER HEADS WERE FOUND TO HAVE CRACKS ON THE UPPER CASING. THE CRACKS WERE LOCATED IN THE SAME AREA OF THE WARMER HEADS, WITH ONE UNIT HAVING DEEPER CRACKS COMPARED TO THE OTHER TWO. IT WAS ALSO SPECIFIED THAT THE HOSPITAL WAS USING INCIDIN FOAM FOR CLEANING AND DISINFECTING THE WARMERS. FURTHERMORE, THEY WERE CLAIMING THAT CLEANING WAS BEING DONE ONLY AFTER THE UNITS HAD BEEN COOLED. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COSYCOT INFANT WARMER FMT FISHER & PAYKEL HEALTHCARE LTD IW931

Patients

Seq Age Sex Outcome Treatment
1