COSYCOT INFANT WARMER
Report
- Report Number
- 9611451-2008-00249
- Event Type
- Malfunction
- Date Received
- May 5, 2008
- Report Date
- August 24, 2007
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- FMT
- PMA / PMN Number
- K971695
- Removal / Correction Number
- Z-0485-2007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADD'L LOT NUMBERS: 050530, 060905. ADD'L MFG DATES: 05/30/2005, 09/05/2006. REPLACE THE ELEVATOR BASES OF THE INFANT WARMERS. METHOD: TWO OUT OF 12 LEANING BASES WERE RETURNED FOR INVESTIGATION. RESULTS: ONE OF THE RETURNED BASES WAS FOUND TO BE LEANING TO THE LEFT. THE OTHER ONE, WHICH WAS ALSO LEANING, HAD CRACKED RIBS AT THE REAR LEFT LEG. LEANING BASES AND CRACKING LEGS OF THE COSYCOT INFANT WARMER, DUE TO EXCESSIVE WEIGHT LOADING, ARE KNOWN PROBLEMS. TOTAL WEIGHT OF THE DEVICE, INCLUDING ACCESSORIES AND PT, EXCEEDING 115 KG MAY, UNDER CERTAIN CIRCUMSTANCES, CAUSE CRACKS IN THE PLASTIC LEGS' BASE AND DAMAGE TO THE BASE ELECTRIC ELEVATOR MECHANISM. CONCLUSION: WE HAVE AN OPEN CAPA TO ADDRESS THESE ISSUES. THE CORRECTIVE ACTION, INFORMING THE USERS OF THE MAXIMUM WEIGHT FOR THE COSYCOT INFANT WARMER AND TO INSPECT THE BASES OF THEIR COSYCOTS, IS EXPECTED TO BE COMPLETED BY END OF MAY 2008. THIS CORRECTIVE ACTION HAS BEEN NOTIFIED TO THE (B)(4) DISTRICT OFFICE UNDER THE REPORTING NUMBER Z-0485-2007.
A HOSPITAL IN (B)(6), REPORTED THAT TWELVE IW930 COSYCOT INFANT WARMERS WERE LEANING TO THE LEFT, 10 OF WHICH HAD UPS (UNINTERRUPTIBLE POWER SUPPLY) ON THE BASES. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COSYCOT INFANT WARMER | FMT | FISHER & PAYKEL HEALTHCARE, LTD. | IW930 | 050513 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |