FDA Adverse Event Injury Summary report: N

AIRLIFE INFANT SKIN TEMPERATURE PROBE, REUSABLE (GIRAFFE, PANDA)

MDR report key: 10382894 · Received August 7, 2020

Report

Report Number
3005934741-2020-00002
Event Type
Injury
Date Received
August 7, 2020
Date of Event
May 2, 2020
Report Date
July 9, 2020
Manufacturer
VYAIRE MEDICAL
Product Code
FMZ
UDI-DI
10885403260759
PMA / PMN Number
K101778
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

RESULT OF INVESTIGATION: THREE PROBES WERE RETURNED WHICH ARE WORKING PROPERLY IN TERMS OF READING TEMPERATURE. HOWEVER, NO INSULATION RESISTANCE OR HIPOT TESTING PERFORMED DUE PRESENCE OF BREAKS IN THE CABLE JACKETS EXPOSING THE SHIELD WIRES. HEAT SHRINK WAS APPLIED ON ONE OF THE PROBES (NOT AS THE ORIGINAL CONDITION) IN ORDER TO KEEP USING IT AFTER THE CABLE JACKET HAD BEEN BREACHED. ALSO, PHOTO INCLUDED IN PACKAGE SHOWS BURN MAY HAVE BEEN CAUSED BY THE TIP OF THE PROBE (NOT THE INCUBATOR). BASED ON THE FINDINGS, IT WAS DETERMINED THAT THE REPORTED FAILURE IS NOT RELATED TO THE MANUFACTURING PROCESS BUT PERHAPS TO A USER ERROR BASED ON THE IFU (INSTRUCTIONS FOR USE) SUPPLIED TO EACH PROBE TO NOT TRY TO USE OR REPAIR A DAMAGED UNIT.

Additional Manufacturer Narrative · 1

AT THIS TIME, VYAIRE HAS NOT RECEIVED THE SUSPECT DEVICE/COMPONENT FOR EVALUATION. THEREFORE, NO ROOT CAUSE COULD BE DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT A PATIENT SUFFERED A BURN INJURY INVOLVING THE AIRLIFE INFANT SKIN TEMPERATURE PROBE, AND MEDICAL INTERVENTION WAS REQUIRED TO TREAT THE WOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
844242 AIRLIFE INFANT SKIN TEMPERATURE PROBE, REUSABLE (GIRAFFE, PANDA) INCUBATOR, NEONATAL FMZ VYAIRE MEDICAL AIRLIFE¿ INFANT SKIN TEMPERATURE PROBE, REUSABLE (GIRAFFE¿, PANDA¿) THM171105 10885403260759

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention