VDW SILVER RECIPROCATING MOTOR
Report
- Report Number
- 9611053-2020-00192
- Event Type
- Malfunction
- Date Received
- August 7, 2020
- Report Date
- August 7, 2020
- Manufacturer
- DENTSPLY VDW GMBH
- Product Code
- EBW
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
AS A RESULT OF THIS MALFUNCTION, THE POTENTIAL FOR SURGICAL INTERVENTION EXISTS TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION AS EVIDENCED BY PREVIOUS REPORTED EVENTS WITH SIMILAR FILES. THIS EVENT, THEREFORE, IS REPORTABLE PER 21CFR PART 803. PLEASE NOTE THAT WHILE THIS PRODUCT IS NOT SOLD IN THE US, IT IS CONSIDERED SIMILAR TO PRODUCTS THAT ARE MARKETED IN THE US BY DENTSPLY (B)(4). BATTERY (WAS CHARGED WITH AN OPERATING TIME OF 18.35.13 HOURS, 360 X, WHEREBY 9 CHARGING CYCLES WOULD HAVE BEEN SUFFICIENT) DEFECTIVE, REPLACED. MICROMOTOR (B)(4), CONTRA-ANGLE 05959 (2007) AND FOOT CONTROL 07073 CHECKED, WITHOUT ERRORS. THE HOUSING HAS CRACKED IN SEVERAL PLACES (PROBABLY DUE TO A FALL) HAS NO EFFECT ON THE FUNCTION. WE, VDW GMBH, DECLARE HEREWITH THAT THE VDW SILVER S/N (B)(4) WITH SOFTWARE (B)(4), MICROMOTOR S/N (B)(4) AND TEST CONTRA ANGLE S/N (B)(4) DOES NOT MEETS THE SPECIFICATION CONCERNING THE TORQUE LIMITATION FUNCTION IN ROTARY MOVEMENT AND RECIPROCATING MOVEMENT. THE TORQUE LIMITATION FUNCTION WAS VERIFIED WITH OUR VDW MOTOR TEST BENCH (P/N 853-093-000-011, S/N (B)(4)).
IN THIS EVENT IT WAS REPORTED THAT A SILVER MOTOR HAS AN INCORRECT TORQUE AND CAUSED FILE BREAKAGE. THE FRAGMENT STILL REMAINS IN A PATIENT'S ROOT CANAL. THE FRAGMENT WILL BE REMOVED AT THE NEXT APPOINTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 845155 | VDW SILVER RECIPROCATING MOTOR | CONTROLLER, FOOT, HANDPIECE AND CORD | EBW | DENTSPLY VDW GMBH | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |