FDA Adverse Event Malfunction Summary report: N

TFCC MENDER DISPOSABLE SUTURE SYSTEM

MDR report key: 10378390 · Received August 7, 2020

Report

Report Number
1219602-2020-01195
Event Type
Malfunction
Date Received
August 7, 2020
Date of Event
July 7, 2020
Report Date
September 2, 2020
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HWQ
UDI-DI
03596010563279
PMA / PMN Number
K885311
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED 7210752 TFCC MENDER DISPOSABLE SUTURE SYSTEM, INTENDED FOR USE IN TREATMENT, HAS NOT RETURNED FOR EVALUATION. WITHOUT THE REPORTED PRODUCT A VISUAL AND FUNCTIONAL EVALUATION CANNOT BE PERFORMED AND THE CUSTOMER¿S COMPLAINT CANNOT BE CONFIRMED. FROM THE INFORMATION PROVIDED, ¿DURING AN ANTERIOR HORN MENISCAL REPAIR, THE NEEDLE OF THE MENDER SUTURE SYSTEMS KEPT BENDING UPON ENTRY INTO THE PATIENT'S JOINT, THIS WOULD NOT ALLOW THE SUTURE TO BE SHUTTLED THROUGH¿. AN EXACT ROOT CAUSE CANNOT BE DETERMINED WITH CONFIDENCE; HOWEVER, FACTORS THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT INCLUDE: EXCESSIVE FORCE APPLIED DURING USE. THE INSTRUCTION FOR USE OUTLINES PRECAUTIONARY STATEMENTS AND INSTRUCTIONS IN REGARDS TO THE USE OF THE DEVICE TO AVOID DAMAGE OR NON-FUNCTIONALITY. THERE WERE NO INDICATIONS THAT WOULD SUGGEST THAT THE DEVICE DID NOT MEET PRODUCT SPECIFICATIONS UPON RELEASE INTO DISTRIBUTION. A REVIEW OF THE COMPLAINT AND MANUFACTURING RECORDS WAS PERFORMED AND INDICATED SIMILAR ALLEGATIONS FOR THE LOT NUMBER REPORTED

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ANTERIOR HORN MENISCAL REPAIR, THE NEEDLE OF THE MENDER SUTURE SYSTEMS KEPT BENDING UPON ENTRY INTO THE PATIENT'S JOINT, THIS WOULD NOT ALLOW THE SUTURE TO BE SHUTTLED THROUGH. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURIES WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
842256 TFCC MENDER DISPOSABLE SUTURE SYSTEM PASSER HWQ SMITH & NEPHEW, INC. 7210752 2034283 03596010563279

Patients

Seq Age Sex Outcome Treatment
1