FDA Adverse Event
Malfunction
Summary report: N
TCM II Y CIRCUIT
MDR report key: 10377
·
Received December 6, 1993
Report
- Report Number
- MW1000016
- Event Type
- Malfunction
- Date Received
- December 6, 1993
- Date of Event
- September 16, 1993
- Report Date
- November 12, 1993
- Manufacturer
- SARNS 3M HEALTH CARE, INC.
- Product Code
- MJJ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
DURING CARDIOPULMONARY BYPASS, FOLLOWING REMOVAL OF THE AORTIC CLAMP, THE H20 TEMPERATURE OF THE SYSTEMIC HEATER/COOLER WAS ALLOWED TO COOL RATHER THAN WARM. THE WATER TEMPERATURE DROPPED FROM 40 DEGREES C TO 26 DEGREES C. THIS CAUSED A DRIP IN BLOOD TEMPERATURE FROM 36 DEGRESS C TO 28 DEGREES C. A DIFFERENT HEATER/COOLER WAS BROUGHT IN TO COMPLETE THE PROCEDURE. NO ADVERSE EFFECT ON PT. A COMBINATION OF VALVES AND TEES RECOMMENDED BY THE MFR MAY CONTRIBUTE TO OPERATOR ERROR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TCM II Y CIRCUIT | 1/2" 2-WAY BALL VALVE | MJJ | SARNS 3M HEALTH CARE, INC. | 14191010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |