FDA Adverse Event Malfunction Summary report: N

TCM II Y CIRCUIT

MDR report key: 10377 · Received December 6, 1993

Report

Report Number
MW1000016
Event Type
Malfunction
Date Received
December 6, 1993
Date of Event
September 16, 1993
Report Date
November 12, 1993
Manufacturer
SARNS 3M HEALTH CARE, INC.
Product Code
MJJ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

DURING CARDIOPULMONARY BYPASS, FOLLOWING REMOVAL OF THE AORTIC CLAMP, THE H20 TEMPERATURE OF THE SYSTEMIC HEATER/COOLER WAS ALLOWED TO COOL RATHER THAN WARM. THE WATER TEMPERATURE DROPPED FROM 40 DEGREES C TO 26 DEGREES C. THIS CAUSED A DRIP IN BLOOD TEMPERATURE FROM 36 DEGRESS C TO 28 DEGREES C. A DIFFERENT HEATER/COOLER WAS BROUGHT IN TO COMPLETE THE PROCEDURE. NO ADVERSE EFFECT ON PT. A COMBINATION OF VALVES AND TEES RECOMMENDED BY THE MFR MAY CONTRIBUTE TO OPERATOR ERROR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TCM II Y CIRCUIT 1/2" 2-WAY BALL VALVE MJJ SARNS 3M HEALTH CARE, INC. 14191010

Patients

Seq Age Sex Outcome Treatment
1 42 YR