FDA Adverse Event Injury Summary report: N

5.0MM FLEXIBLE SHAFT

MDR report key: 10376331 · Received August 6, 2020

Report

Report Number
2939274-2020-03418
Event Type
Injury
Date Received
August 6, 2020
Date of Event
July 16, 2020
Report Date
July 16, 2020
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HTO
UDI-DI
10886982194114
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: H3, H6: INVESTIGATION SUMMARY INVESTIGATION FLOW: DAMAGE VISUAL INSPECTION: THE 5.0MM FLEXIBLE SHAFT (PART # 352.040 AND LOT #: 2201779) WAS RECEIVED AT US CQ. UPON THE VISUAL INSPECTION OF THE COMPLAINT DEVICE, IT WAS SEEN TO BE BROKE TOWARDS THE DISTAL FLANGES AND THE BROKEN FLANGES WERE RETURNED. THE EMBEDDED PIECE CANNOT BE CONFIRMED AS THERE WERE NO PHOTOS/X-RAYS PROVIDED . NO OTHER ISSUES WERE IDENTIFIED WITH THE RETURNED DEVICE. DEVICE FAILURE/DEFECT IDENTIFIED? YES DIMENSIONAL INSPECTION: A DIMENSIONAL INSPECTION WAS NOT PERFORMED DUE TO POST MANUFACTURING DAMAGE. DOCUMENT/SPECIFICATION REVIEW: NO DESIGN ISSUES OR DISCREPANCIES WERE IDENTIFIED. COMPLAINT CONFIRMED? YES INVESTIGATION CONCLUSION: THIS COMPLAINT CONDITION IS CONFIRMED SINCE THE SHAFT WAS BROKEN TOWARDS THE DISTAL FLANGES. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED BASED ON THE PROVIDED INFORMATION. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT PART NUMBER: 352.040 SYNTHES LOT NUMBER: 2201779 MANUFACTURING SITE: BETTLACH RELEASE TO WAREHOUSE DATE: JUNE 23, 2006 A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON (B)(6) 2020, DURING AN UNKNOWN PROCEDURE, THE 5.MM FLEXIBLE SHAFT AND THE 8.5MM MEDULLARY REAMER HEAD WERE BROKEN IN THE PATIENT. THE MAJORITY OF THE PIECES WERE RETRIEVED, HOWEVER, THERE WERE ONE TO TWO PIECES THAT WERE STILL LOCKED IN THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. THE PATIENT OUTCOME WAS SUCCESSFUL. THIS REPORT IS FOR 1 5.0MM FLEXIBLE SHAFT. THIS IS REPORT 1 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
836136 5.0MM FLEXIBLE SHAFT REAMER HTO WRIGHTS LANE SYNTHES USA PRODUCTS LLC 352.040 2201779 10886982194114

Patients

Seq Age Sex Outcome Treatment
1 UNK - REAMERS: REAMER HEAD.