FDA Adverse Event Malfunction Summary report: N

ELITE IQ

MDR report key: 10376132 · Received August 6, 2020

Report

Report Number
1222993-2020-00019
Event Type
Malfunction
Date Received
August 6, 2020
Date of Event
July 22, 2020
Report Date
August 6, 2020
Manufacturer
EL. EN. ELECTRONIC ENGINEERING SPA
Product Code
GEX
Product Problem
Yes
Report Source
Distributor report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ELITE IQ WAS UNABLE TO ACHIEVE DESIRED FLUENCES. AN INVESTIGATION WAS CONDUCTED ON 7/22 AND FOUND ENERGY EMITTED WAS LOW AND MEASUREMENT SHOWED IT WAS GREATER THAN 20% OUT OF SPECIFICATION. THERE WAS NO REPORT OF A PATIENT INJURY RELATED TO THIS EVENT. THIS INCIDENT IS REPORTABLE BECAUSE IT IS AN ACCIDENTAL RADIATION OCCURRENCE (ARO) SINCE THE DEVICE OPERATED OUT OF SPECIFICATION RANGE.