FDA Adverse Event
Malfunction
Summary report: N
ELITE IQ
MDR report key: 10376132
·
Received August 6, 2020
Report
- Report Number
- 1222993-2020-00019
- Event Type
- Malfunction
- Date Received
- August 6, 2020
- Date of Event
- July 22, 2020
- Report Date
- August 6, 2020
- Manufacturer
- EL. EN. ELECTRONIC ENGINEERING SPA
- Product Code
- GEX
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ELITE IQ WAS UNABLE TO ACHIEVE DESIRED FLUENCES. AN INVESTIGATION WAS CONDUCTED ON 7/22 AND FOUND ENERGY EMITTED WAS LOW AND MEASUREMENT SHOWED IT WAS GREATER THAN 20% OUT OF SPECIFICATION. THERE WAS NO REPORT OF A PATIENT INJURY RELATED TO THIS EVENT. THIS INCIDENT IS REPORTABLE BECAUSE IT IS AN ACCIDENTAL RADIATION OCCURRENCE (ARO) SINCE THE DEVICE OPERATED OUT OF SPECIFICATION RANGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 837674 | ELITE IQ | ELITE IQ | GEX | EL. EN. ELECTRONIC ENGINEERING SPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |