ACCUTREND PLUS GLUCOSE STRIPS
Report
- Report Number
- 1823260-2020-01905
- Event Type
- Malfunction
- Date Received
- August 6, 2020
- Date of Event
- July 6, 2020
- Report Date
- August 6, 2020
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- CGA
- PMA / PMN Number
- K051376
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- 003
Narratives
THIS EVENT OCCURRED IN (B)(6). THE CUSTOMER'S METERS AND STRIPS WERE REQUESTED FOR INVESTIGATION. THE PRODUCTS HAVE NOT BEEN RECEIVED AT THIS TIME. RETENTION SAMPLES FOR GLUCOSE LOT 349029-04 WERE MEASURED WITH RETENTION CONTROLS GKL60 LOT 346092-99 /GKL180 LOT 346093-99 AND 3 EDTA BLOOD ON REFERENCE DEVICES COMPARED TO LABORATORY METHOD. RANGES CONTROL SOLUTIONS: RANGE GKL 60: 58 ¿ 98 MG/DL. INVESTIGATION RESULTS USING RETENTION MATERIAL: QC 1: 78 MG/DL, QC 2: 74 MG/DL, QC 3: 72 MG/DL. INVESTIGATION RESULTS USING RETENTION MATERIAL: RANGE GKL 180: 166 ¿ 224 MG/DL, QC 1: 195 MG/DL, QC 2: 194 MG/DL, QC 3: 187 MG/DL. ALL RESULTS WERE WITHIN THE CONTROL RANGE. NO SIGNIFICANT DIFFERENCES TO THE LABORATORY METHOD. THE INVESTIGATION IS ONGOING.
THE INITIAL REPORTER RECEIVED QUESTIONABLE GLUCOSE RESULTS USING AN ACCUTREND PLUS METER WHEN COMPARED TO AN ACCU-CHEK ACTIVE METER. THE METER RESULT ON THE ACCUTREND METER WAS 82 MG/DL. A FEW MINUTES LATER, THE CUSTOMER TESTED ON THE ACCU-CHEK ACTIVE METER. THE RESULT WAS 110 MG/DL. THE CUSTOMER PROVIDED TWO ACCUTREND METER SERIAL NUMBERS (B)(4). IT IS UNKNOWN WHICH METER GAVE THE RESULT. THE ACCU-CHEK ACTIVE METER'S SERIAL NUMBER WAS REQUESTED BUT NOT PROVIDED. IT IS NOT KNOWN WHICH RESULT WAS BELIEVED TO BE CORRECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 838850 | ACCUTREND PLUS GLUCOSE STRIPS | BLOOD GLUCOSE MONITORING TEST STRIPS | CGA | ROCHE DIAGNOSTICS | NA | 34902904 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |